The clinical benefit of cardiac resynchronization therapy optimization using a device-based hemodynamic sensor in a patient with dilated cardiomyopathy: a case report
نویسندگان
چکیده
INTRODUCTION Results on the evolution of the clinical status of patients undergoing cardiac resynchronization therapy with a defibrillator after automatic optimization of their cardiac resynchronization therapy are scarce. We observed a rapid and important change in the clinical status of our non-responding patient following activation of a sensor capable of weekly atrioventricular and interventricular delays' optimization. CASE PRESENTATION A 78-year-old Caucasian man presented with dilated cardiomyopathy, left bundle branch block, a left ventricular ejection fraction of 35 %, New York Heart Association class III/IV heart failure, and paroxysmal atrial fibrillation. Our patient was implanted with a cardiac resynchronization device with a defibrillator and the SonRtip atrial lead. Right ventricular and left ventricular leads were also implanted. Because of the recurrence of atrial fibrillation, the automatic optimization was set off at discharge. Consequently, the device did not optimize atrioventricular and interventricular delays (programming at discharge: 125 ms for the atrioventricular delay and 0 ms for the interventriculardelay). Our patient was treated with an anti-arrhythmic drug. Five months after implantation, his clinical status remained impaired (left ventricular ejection fraction = 30 %). The SonR signal amplitude had also decreased from 0.52 g to 0.29 g. Nevertheless, because our patient was no longer presenting with atrial fibrillation, the anti-arrhythmic treatment was stopped and the SonR optimization system was activated. After 2 months of automatic cardiac resynchronization therapy with defibrillator optimization, our patient's clinical status had significantly improved (left ventricular ejection fraction = 60 %, New York Heart Association class II) and the SonR signal amplitude had doubled shortly after the first weekly automatic optimization. CONCLUSION In this non-responding patient, device-based automatic cardiac resynchronization therapy optimization was shown to significantly improve his clinical status.
منابع مشابه
Apical Hypertrophic Cardiomyopathy in a Case with Chest Pain and Family History of Sudden Cardiac Death: A Case Report
Hypertrophic cardiomyopathy (HCM) is the most common genetic cardiovascular disease, which is caused by a multitude of mutations in genes encoding proteins of the cardiac sarcomere (1). Apical hypertrophic cardiomyopathy (AHCM) is an uncommon type of HCM. The sudden cardiac death is less likely to occur in the patients inflicted with AHCM (2). Herein, we presented the case of a 29-year-old man ...
متن کاملAdaptive cardiac resynchronization therapy for dilated cardiomyopathy with functional mitral regurgitation
We report the case of a man in his 60s who had dilated cardiomyopathy with severe functional mitral regurgitation. Four years after a cardiac resynchronization therapy (CRT) device with an implantable cardioverter defibrillator was implanted, this device was replaced with an adaptive CRT device because of battery consumption. Seven months after replacement of this device, the left ventricular p...
متن کاملAnesthetic management of patients with dilated cardiomyopathy for noncardiac surgery.
Anesthetic management of patients with dilated cardiomyopathy (DCM) is a challenge to the anesthesiologist, due to poor left systolic function, ventricular enlargement, risk of malignant arrhythmias and sudden cardiac death. Therefore, preoperative assessment and appropriate anesthetic management are important in patients with DCM. This review describes the preoperative evaluation and anesthesi...
متن کاملEvaluation of Left Ventricular Endocardial Cardiac Resynchronization Therapy in a Non-responder with Ventricular Arrhythmias
Approximately one third of patients treated with cardiac resynchronization therapy do not derive any detectable benefit. In these patients, acute invasive hemodynamic evaluation can be used for therapy optimization. This report describes the use of systematic invasive hemodynamic measurements for clinical decision making in a patient who experienced severe ventricular arrhythmias and clinical d...
متن کاملCardiac resynchronization therapy improves the uptake of MIBI-99mTc and cardiac function.
This case shows the improvement promoted by cardiac resynchronization therapy (CRT) on myocardial perfusion and left ventricular (LV) performance assessed by gated myocardial perfusion scintigraphy. The patient had idiopathic dilated cardiomyopathy, left bundle branch block and severe heart failure despite optimized medical treatment. After CRT, clinical improvement, QRS reduction and improveme...
متن کامل