Charge Collection Characteristics of Frisch Collar CdZnTe Gamma Ray Spectrometers
نویسندگان
چکیده
A collimated Au source was used to determine the charge collection efficiency (CCE) at several locations along the length of a 3.4 x 3.4 x 5.5 mm CdZnTe bar detector both in planar configuration and with Frisch collars of varying length. For each configuration, a 0.50 mm long region spanning the width of the device was irradiated with 411 keV gamma rays produced by a neutron activated gold foil. Irradiation began at the cathode and stepped in 0.50 mm steps toward the anode with a spectrum being collected at each location. By observing the channel location of the full energy peak in each collected spectrum, an average CCE was determined for each irradiated region. The CCE was found to vary nearly linearly along the length of the device in the planar configuration, starting at a peak value of 89% and dropping to a minimum measured value of 26% near the anode. The addition of a Frisch collar covering the entire length of the crystal greatly altered the CCE profile, which remained near 87% for approximately two-thirds of the length, then sharply dropped near the anode. Results were confirmed by theoretical models. Further CCE mapping was also completed for devices with Frisch collars of various lengths. Those results are report as well. © 2001 Elsevier Science. All rights reserved
منابع مشابه
Angular response of a W-collimated room-temperature-operated CdZnTe Frisch collar spectrometer
The spatial resolution of a collimated CdZnTe Frisch collar detector was experimentally investigated for two different tungsten collimator designs using a Au gamma-ray source at 412 keV. A two-dimensional model for the detector–collimator–source geometry was developed and applied with some simplifying assumptions. The CdZnTe detector was fabricated from a single crystal using a copper shim as t...
متن کاملGeometrically weighted semiconductor Frisch grid radiation spectrometers
A new detector geometry is described with relatively high gamma ray energy resolution at room temperature. The device uses the geometric weighting effect, the small pixel effect and the Frisch grid effect to produce high gamma ray energy resolution. The design is simple and easy to construct. The device performs as a gamma ray spectrometer without the need for pulse shape rejection or correctio...
متن کاملCdZnTe Frisch collar detectors for g-ray spectroscopy
Low-energy g-ray spectra were collected from Am, Co, Ba, Au, Cs and U using a 3.4 3.4 5.7mm CdZnTe detector utilizing an insulated Frisch ring. The CdZnTe detector was fabricated from a single crystal and a copper shim was used as the Frisch collar. Room-temperature energy resolution of 1.45% full-width half-maximum (FWHM) was obtained for Cs at 661.7 keV without electronic correction. The dete...
متن کاملE!ects of charge sharing in 3-D position sensitive CdZnTe gamma-ray spectrometers
Two 3-D position-sensitive semiconductor c-ray spectrometers operable at room temperature, have been fabricated using 1 cm3 cubic CdZnTe crystals. c-Ray spectra have been obtained from 11(x)]11(y)]20(z) voxels, each corresponding to &0.7]0.7]0.5 mm in the detector volume. Individual spectra from each of the 2420 voxels were corrected for electron trapping, di!erences in weighting potentials, an...
متن کاملA high pressure xenon gamma-ray spectrometer using a coplanar anode configuration
A new design of a high pressure xenon ionization chamber has been fabricated in an attempt to eliminate the problems associated with acoustical vibrations of the Frisch grid. The function of the traditional Frisch grid has been accomplished by employing a coplanar anode system capable of single polarity charge sensing. Two different detector designs have been fabricated using both cylindrical a...
متن کامل