Detection of very delayed coronary stent fracture using a novel fluoroscopic stent visualization technique against intravascular ultrasound
نویسندگان
چکیده
RATIONALE Stent fracture has received increased concern as it may be an important risk factor for late stent failure, intravascular ultrasound (IVUS) is always recommended to confirm the diagnosis of stent fracture. StentBoost can detect stent fractures more easily due to the enhanced stent strut visibility, compared with coronary angiography (CAG). Few cases were reported to compare the advantages of StentBoost vis-à-vis IVUS in detecting stent fracture. PATIENT CONCERNS We reported 3 cases that were confirmed the diagnosis of stent fracture by StentBoost, which were preliminarily suspected by angiography, including one case that lacked the IVUS evidence of stent fracture. DIAGNOSES 3 cases of stent fracture presented with asymptomatic, angina and acute myocardial infarction, respectively. INTERVENTION Stents were implanted in the patients of case 2 and case 3, but the patient of case 1 was not given any intervention. OUTCOMES No recurrent angina or myocardial infarction during outpatient follow-up. LESSONS StentBoost may distinguish partial, complete, or multiple stent fracture, even which sometimes is not obvious in IVUS, StentBoost is a useful and handy tool for identifying the stent struts.
منابع مشابه
Optimal Stent Expansion by Stent Balloon Multiple Inflation at Nominal Pressure in Resistant Lesions: A New Technique
Introduction: This study aimed to evaluate the effects of four-time inflation of the stent balloon at nominal pressure on optimal stent expansion in resistant lesions. Materials and Methods: This interventional study was conducted on 39 patients with coronary artery lesions, in whom Zotarolimus-eluting stents (N=20), Paclitaxel-eluting stents (N=11) and other stents (N=8) were deployed four tim...
متن کاملStentBoost Visualization for the Evaluation of Coronary Stent Expansion During Percutaneous Coronary Interventions
INTRODUCTION Inadequate stent implantation is associated with stent thrombosis and restenosis. StentBoost can enhance stent visualization and evaluate stent expansion. Currently, there are limited comparison studies between StentBoost and intravascular ultrasound (IVUS). We aimed to test the correlation and agreement between IVUS and StentBoost measurements. METHODS From December 2010 to Dece...
متن کاملIntra-coronary Stent Localization in Intravascular Ultrasound Sequences, A Preliminary Study
An intraluminal coronary stent is a metal scaffold deployed in a stenotic artery during Percutaneous Coronary Intervention (PCI). Intravascular Ultrasound (IVUS) is a catheter-based imaging technique generally used for assessing the correct placement of the stent. All the approaches proposed so far for the stent analysis only focused on the struts detection, while this paper proposes a novel ap...
متن کاملStent Evaluation with Optical Coherence Tomography
Optical coherence tomography (OCT) has been recently applied to investigate coronary artery disease in interventional cardiology. Compared to intravascular ultrasound, OCT is able to visualize various vascular structures more clearly with higher resolution. Several validation studies have shown that OCT is more accurate in evaluating neointimal tissue after coronary stent implantation than intr...
متن کاملManaging patients with intermediate in-stent restenotic lesions: is it "prime time" for intravascular ultrasound imaging?
The importance of intravascular ultrasound (IVUS) as an adjunctive technology facilitating optimal coronary interventional procedural technique has become increasingly evident over the past decade. IVUS is currently appreciated as an invaluable tool enabling better assessment of reference vessel size and coronary lesion morphology, including gross plaque characteristics and lesion length. Accor...
متن کامل