Evaluation of different disinfectants on the performance of an on-meter dosed amperometric glucose-oxidase-based glucose meter.

نویسندگان

  • Don Sarmaga
  • Jeffrey A Dubois
  • Martha E Lyon
چکیده

BACKGROUND Off-meter dosed photometric glucose-oxidase-based glucose meters have been reported to be susceptible to interference by hydrogen-peroxide-based disinfecting agents. The objective of this study was to determine if a single application of hydrogen-peroxide-containing Accel® wipe to disinfect an on-meter dosed amperometric glucose-oxidase-based glucose meter will influence its performance. METHOD The performance of five on-meter dosed amperometric glucose-oxidase-based glucose meters was determined before and after disinfecting the devices with a single application of either CaviWipes® (14.3% isopropanol and 0.23% diisobutyl-phenoxy-ethoxyethyl dimethyl benzyl ammonium chloride) or Accel (0.5% hydrogen peroxide) wipes. Replicate glucose measurements were conducted before disinfecting the devices, immediately after disinfecting, and then 1 and 2 min postdisinfecting, with measurements in triplicate. Analysis was sequentially completed for five different meters. Results were analyzed by a two-way analysis of variance (Analyze-it software). RESULTS No clinical (<0.3 mmol/liter) or statistical differences (p > .05) in glucose concentration were detected when the on-meter dosed amperometric glucose-oxidase-based glucose meters were disinfected with either CaviWipes or Accel wipes and measured immediately or 1 or 2 min postdisinfecting. No clinically significant difference in glucose concentration was detected between meters (<0.3 mmol/liter). CONCLUSION The on-meter dosed glucose oxidase amperometric-based glucose meters are not analytically susceptible to interference by a single application of hydrogen-peroxide-containing Accel disinfectant wipes.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Inkjet Printing of Reducible AgNPs amperometric glucose biosensor Electrodes

The enzymes immobilization of the is crucially effective factor in biosensor preparation. Metal nanoparticles potentially able to immobilize the enzymes according to unique properties including large surface-to-volume ratio, high surface reaction activity, high catalytic efficiency, and strong adsorption ability. A novel and highly sensitive amperometric glucose biosensor was obtained by using ...

متن کامل

Evaluation of a point-of-care electrochemical meter to detect subclinical ketosis and hypoglycaemia in lactating dairy cows.

OBJECTIVES To evaluate and validate a hand-held electrochemical meter (Precision Xtra®) as a screening test for subclinical ketosis and hypoglycaemia in lactating dairy cattle. DESIGN Method comparison study using a convenience sample. PROCEDURE Blood samples were collected into plain tubes from the coccygeal vessels of 181 Holstein cows at 2-4 weeks of lactation during summer in Iran. Bloo...

متن کامل

Analysis Article on the Performance Analysis of the OneTouch® UltraVueTM Blood Glucose Monitoring System

In this issue of Journal of Diabetes Science and Technology, Chang and colleagues present the analytical performance evaluation of the OneTouch® UltraVueTM blood glucose meter. This device is an advanced construction with a color display, used-strip ejector, no-button interface, and short assay time. Accuracy studies were performed using a YSI 2300 analyzer, considered the reference. Altogether...

متن کامل

Comparison of Two Veterinary Blood Glucose Meters and One Human-Based Glucose Meter for Use in Dogs

Background: Recently, tendency to use veterinary specific Portable blood glucose meter (PBGMs) has increased. However, assessment of their analytical and clinical accuracy is a matter of concern. OBJECTIVES: To assess accuracy of two veterinary (AlphaTRAK2 and CERA-PET) and one human-based (Bionime) PBGMs for canine blood samples. METHODS: In this cross-sectional study, a total of 160 client-ow...

متن کامل

Analytical Performance and Quality Control of a Glucose Monitor System

  Background and Objective: The reliability and validity of monitors for self-monitoring of blood glucose are debated. We evaluated the analytical performance of Accu-check Active (Boehringer Mannheim, Roche) which is one of the most commonly used monitors in Iran. Material and Methods: We compared the monitor readings with the reference values by percentage of values within...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of diabetes science and technology

دوره 5 6  شماره 

صفحات  -

تاریخ انتشار 2011