Effects of multiple photon scattering on maximum likelihood positioning in PET
نویسندگان
چکیده
Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen, Germany Conventional positioning algorithms in PET assume a photon interaction and complete energy deposition in one detector element (single-hit). However, there is a high probability that a photon interacts within a crystal element via Compton scattering but is absorbed in another element (multiple-hit). Thus scintillation light is generated in multiple crystals and the resulting light distribution is spread over more detector elements. This leads, among others, to a degradation of the spatial resolution, especially in pixelated scintillators with a small pitch, unless more sophisticated positioning algorithms are exploited. In this work, we use a maximum-likelihood method (MLM) to identify the 2D hit position with a single-hit probability model. We want to study the characteristics of multiple-hit events, to implement a rejection filter using a likelihood threshold, which will consequently be beneficial for the intrinsic spatial resolution. For the studies we use a sensor tile with 8×8 digital SiPMs. A 30×30 LYSO crystal array (crystal volume 1×1×12mm3) is mounted on top of the tile using an intermediate light guide (2mm). This setup was used to measure the single-hit probability density for the MLM using a constrained center-of-gravity positioning. To study the MLM, we simulate the different event classes using the above described setup with an additional 511keV point source. The response of the SiPMs to a simulated event is modeled in two steps. First the deposited energy is determined for each hit crystal with intrinsic energy resolution and converted into photon count. Second this photon count is spread over the SiPMs with the related mean of the measured probability density. For multiple-hit events, the calculated photon count of all hit crystals for each SiPM is summed up. We are able to demonstrate, that a likelihood threshold can substantially improve the detector intrinsic spatial resolution of blurring caused by multiple-hit events. Results and details about this study will be presented at the conference.
منابع مشابه
Effects of multiple-interaction photon events in a high-resolution PET system that uses 3-D positioning detectors.
PURPOSE The authors' laboratory is developing a dual-panel, breast-dedicated PET system. The detector panels are built from dual-LSO-position-sensitive avalanche photodiode (PSAPD) modules-units holding two 8 x 8 arrays of 1 mm3 LSO crystals, where each array is coupled to a PSAPD. When stacked to form an imaging volume, these modules are capable of recording the 3-D coordinates of individual i...
متن کاملComparative assessment of the accuracy of maximum likelihood and correlated signal enhancement algorithm positioning methods in gamma camera with large square photomultiplier tubes
Introduction: The gamma cameras, based on scintillation crystal followed by an array of photomultiplier tubes (PMTs), play a crucial role in nuclear medicine. The use of square PMTs provides the minimum dead zones in the camera. The camera with square PMTs also reduces the number of PMTs relative to the detection area. Introduction of a positioning algorithm to improve the spat...
متن کاملروشی برای شبیه سازی ترابرد فوتون درماده
Monte Carlo simulation is widely used in calculations involing transport of photons through different materials of different shapes. The method consists of randomly generating a finite set of photon histories over which the quantities of interest are averaged. In photon transport calculations, sampling the photon scattering angle from the Klein-Nishina probability distribution is of special i...
متن کاملBayesian reconstruction of photon interaction sequences for high-resolution PET detectors.
Realizing the full potential of high-resolution positron emission tomography (PET) systems involves accurately positioning events in which the annihilation photon deposits all its energy across multiple detector elements. Reconstructing the complete sequence of interactions of each photon provides a reliable way to select the earliest interaction because it ensures that all the interactions are...
متن کاملApplication of maximum likelihood methods to laser Thomson scattering measurements of low density plasmas.
Laser Thomson scattering (LTS) is an established plasma diagnostic technique that has seen recent application to low density plasmas. It is difficult to perform LTS measurements when the scattered signal is weak as a result of low electron number density, poor optical access to the plasma, or both. Photon counting methods are often implemented in order to perform measurements in these low signa...