چکیده
An optical surgical navigation system is used as a navigator to facilitate surgical approaches, and pulse oximeters provide valuable information for anesthetic management. However, saw-tooth waves on the monitor of a pulse oximeter and the inability of the pulse oximeter to accurately record the saturation of a percutaneous artery were observed when a surgeon started an optical navigation system. The current case is thought to be the first report of this navigation system interfering with pulse oximetry. The causes of pulse jamming and how to manage an optical navigation system are discussed.
منابع مشابه
Zero Oxygen Saturation in Pulse Oximeter: Need for Clinical Correlation
Sir, I would like to share an unusual incident, which had happened recently in our operation theater. This event had occurred while conducting spinal anesthesia for a 28-year-male patient with an inguinal hernia for meshplasty. His preoperative anesthetic evaluation and investigations were unremarkable. In the operating room, the blood pressure (BP) cuff and pulse oximeter were applied in the s...
متن کاملInterference with pulse oximetry by the Stealth Station™ Image Guidance System
Background A pulse oximeter is one of the most important monitors to save patients undergoing anesthesia and monitored sedation. The authors report a case of orthopedic surgery, in which interference of pulse oximetry occurred when using a Stealth Station™ navigation system (Medtronic Sofamor Danek, Memphis, TN). Applying a black plastic shield (Masimo Ambient Shield: Masimo Corporation, Irvine...
متن کاملCapstone Design 14:332:468 A Wearable Pulse Oximeter
In this report, the design and implementation of a low-cost, portable and wearable pulse oximeter is presented. A pulse oximeter is a non-invasive device capable of monitoring the blood’s oxygen saturation. It has been widely used in the medical, fitness and clinical care worlds. A low-cost wearable oximeter can significantly expand its applicability. The goal of this capstone design project wa...
متن کاملNovel Wearable and Wireless Ring-Type Pulse Oximeter with Multi-Detectors
The pulse oximeter is a popular instrument to monitor the arterial oxygen saturation (SPO2). Although a fingertip-type pulse oximeter is the mainstream one on the market at present, it is still inconvenient for long-term monitoring, in particular, with respect to motion. Therefore, the development of a wearable pulse oximeter, such as a finger base-type pulse oximeter, can effectively solve the...
متن کاملEvaluation of the Ohmeda 3700 pulse oximeter.
Arterial oxygen saturation values (Sao2) from 60% to 98% were measured by the Ohmeda 3700 pulse oximeter with the three types of probe available and compared with values of oxygen saturation estimated from direct arterial sampling (arterial oxygen and carbon dioxide tensions and pH) on 65 occasions. The response time of the oximeter was measured after a sudden rise in inspired oxygen concentrat...