Beam modeling and VMAT performance with the Agility 160‐leaf multileaf collimator
نویسندگان
چکیده
The Agility multileaf collimator (Elekta AB, Stockholm, Sweden) has 160 leaves of projected width 0.5 cm at the isocenter, with maximum leaf speed 3.5 cms-1. These characteristics promise to facilitate fast and accurate delivery of radiotherapy, particularly volumetric-modulated arc therapy (VMAT). The aim of this study is therefore to create a beam model for the Pinnacle3 treatment planning system (Philips Radiation Oncology Systems, Fitchburg, WI), and to use this beam model to explore the performance of the Agility MLC in delivery of VMAT. A 6 MV beam model was created and verified by measuring doses under irregularly shaped fields. VMAT treatment plans for five typical head-and-neck patients were created using the beam model and delivered using both binned and continuously variable dose rate (CVDR). Results were compared with those for an MLCi unit without CVDR. The beam model has similar parameters to those of an MLCi model, with interleaf leakage of only 0.2%. The verification of irregular fields shows a mean agreement between measured and planned dose of 1.3% (planned dose higher). The Agility VMAT head-and-neck plans show equivalent plan quality and delivery accuracy to those for an MLCi unit, with 95% of verification measurements within 3% and 3 mm of planned dose. Mean delivery time is 133 s with the Agility head and CVDR, 171 s without CVDR, and 282 s with an MLCi unit. Pinnacle3 has therefore been shown to model the Agility MLC accurately, and to provide accurate VMAT treatment plans which can be delivered significantly faster with Agility than with an MLCi.
منابع مشابه
An experimental evaluation of the Agility MLC for motion-compensated VMAT delivery.
An algorithm for dynamic multileaf-collimator (dMLC) tracking of a target performing a known a priori, rigid-body motion during volumetric modulated arc therapy (VMAT), has been experimentally validated and applied to investigate the potential of the Agility (Elekta AB, Stockholm, Sweden) multileaf-collimator (MLC) for use in motion-compensated VMAT delivery. For five VMAT patients, dosimetric ...
متن کاملFeasibility of single-isocenter, multi-arc non-coplanar volumetric modulated arc therapy for multiple brain tumors using a linear accelerator with a 160-leaf multileaf collimator: a phantom study
The feasibility of single isocenter, multi-arc non-coplanar volumetric modulated arc therapy (VMAT) for multiple brain tumors was studied using an Elekta Synergy linear accelerator with an Agility multileaf collimator and a Monaco treatment planning system. Two VMAT radiosurgery plans consisting of a full arc and three half arcs were created with a prescribed dose of 20 Gy in a single fraction....
متن کاملBeam characteristic of P.E collimators Add-on multileaf collimator
Introduction: Multileaf collimators (MLCs) are computer-controlled devices which are Useful and cost-effective tools for conformal therapy and intensity modulation radiotherapy (IMRT). Nowadays MLCs are important and standard configuration of the new medical linear accelerators which are substitute of conventional Blocks and has eliminated the ...
متن کاملMonte Carlo Study of Unflattened Photon Beams Shaped by Multileaf Collimator
Introduction: This study investigates basic dosimetric properties of unflattened 6 MV photon beam shaped by multileaf collimator and compares them with those of flattened beams.Materials and Methods: Monte Carlo simulation model using BEAM code was developed for a 6MV photon beam based on Varian Clinic 600 unique performance linac operated with and without a flattening filter in beam line. Dosi...
متن کاملImpact of MLC properties and IMRT technique in meningioma and head-and-neck treatments
PURPOSE The impact of multileaf collimator (MLC) design and IMRT technique on plan quality and delivery improvements for head-and-neck and meningioma patients is compared in a planning study. MATERIAL AND METHODS Ten previously treated patients (5 head-and-neck, 5 meningioma) were re-planned for step-and-shoot IMRT (ssIMRT), sliding window IMRT (dMLC) and VMAT using the MLCi2 without (-) and ...
متن کامل