Large Diameter Femoral Heads Impose Significant Alterations on the Strains Developed on Femoral Component and Bone: A Finite Element Analysis
نویسندگان
چکیده
Total Hip Arthroplasty aims at fully recreating a functional hip joint. Over the past years modular implant systems have become common practice and are widely used, due to the surgical options they provide. In addition Big Femoral Heads have also been implemented in the process, providing more flexibility for the surgeon. The current study aims at investigating the effects that femoral heads of bigger diameter may impose on the mechanical behavior of the bone-implant assembly. Using data acquired by Computed Tomographies and a Coordinate Measurement Machine, a cadaveric femur and a Profemur-E modular stem were fully digitized, leading to a three dimensional finite element model in ANSYS Workbench. Strains and stresses were then calculated, focusing on areas of clinical interest, based on Gruen zones: the calcar and the corresponding below the greater trochanter area in the proximal femur, the stem tip region and a profile line along linea aspera. The performed finite elements analysis revealed that the use of large diameter heads produces significant changes in strain development within the bone volume, especially in the lateral side. The application of Frost's law in bone remodeling, validated the hypothesis that for all diameters normal bone growth occurs. However, in the calcar area lower strain values were recorded, when comparing with the reference model featuring a 28mm femoral head. Along line aspera and for the stem tip area, higher values were recorded. Finally, stresses calculated on the modular neck revealed increased values, but without reaching the yield strength of the titanium alloy used.
منابع مشابه
The axisymmetric computational study of a femoral component to analysis the effect of titanium alloy and diameter variation.
This work presents a numerical approach in order to predict the influence of implant material stiffness in a femoral component design when submitted in compression. The implant success depends on the transferred load to the neighboring bone. The finite element method can be used to analysis the stress and strain distribution in the femoral component allowing to improve the implant selection. Fo...
متن کاملAn investigation of the effects of osteoporosis, impact intensity and orientation on human femur injuries: a parametric finite element study
Objective: Femur is the strongest, longest and heaviest bone in the human body. Due to the great importance of femur in human body, its injury may cause large numbers of disabilities and mortality. Considering various effective parameters such as mechanical properties, geometry, loading configuration, etc. can propel the study to the trustable results.. Methods: A 3D finite element model of the...
متن کاملStructural Efficacy of a Novel Porous Tantalum Implant for Osteonecrosis Grafting
Introduction: Vascularized fibular grafting is a promising treatment for femoral head osteonecrosis (ON), the rationale in part being to support the subchondral plate during structural reconstitution of the underlying (weakened) cancellous bone. Unfortunately, donor site morbidity is appreciable, including weakness, discomfort, and sensory abnormalities. An implant made from porous tantalum (He...
متن کاملThe Effect of Sagittal Femoral Bowing on the Femoral Component Position in Total Knee Arthroplasty
Background: In the current study, we investigated that how sagittal femoral bowing can affect the sagittal alignment ofthe femoral component in total knee arthroplasty (TKA).Methods: There were 25 patients underwent TKA. Long leg radiography in lateral view was performed. The sagittalfemoral bowing (SFB) and component alignment in relation to the sagittal mechanical axis and distal anterior cor...
متن کاملFinite element analysis of the resurfaced femoral head.
Failure of the resurfaced femoral head may occur in the short term owing to femoral neck fracture or in the long term owing to aseptic loosening as a result of strain shielding. Resurfacing arthroplasties are not all the same. In particular, there is considerable debate regarding the role of the metaphyseal stem and cementing technique. This study examines the influence of various metaphyseal s...
متن کامل