Intensities of Incident and Transmitted Ultraviolet-A Rays through Gafchromic Films
نویسندگان
چکیده
Gafchromic films have been applied to X-ray dosimetry in diagnostic radiology. To correct nonuniformity errors in Gafchromic films, X-rays in the double-exposure technique can be replaced with ultraviolet (UV)-A rays. Intensities of the incident and transmitted UV-A rays were measured. However, it is unclear whether the chemical color change of Gafchromic films affects the UV-A transmission intensity. Gafchromic EBT3 films were suitable to be used in this study because non-UV protection layers are present on both sides of the film. The film is placed between UV-A ray light-emitting diodes and a probe of a UV meter. Gafchromic EBT3 films were irradiated by UV-A rays for up to 60 min. Data for analysis were obtained in the subsequent 60 min. Images from before and after UV-A irradiation were subtracted. When using 375 nm UV-A, the mean ± standard deviation (SD) of the pixel values in the subtracted image was remarkably high (11,194.15 ± 586.63). However, the UV-A transmissivity remained constant throughout the 60 min irradiation period. The mean ± SD UV-A transmission intensity was 184.48 ± 0.50 μm/cm2. Our findings demonstrate that color density changes in Gafchromic EBT3 films do not affect their UV-A transmission. Therefore, Gafchromic films were irradiated by UV-A rays as a preexposure.
منابع مشابه
Ultraviolet exposure of Gafchromic XR‐RV3 and XR‐SP2 films
Gafchromic film has been used for X-ray dose measurement in diagnostic examinations. Their use has been initiated for three-dimensional X-ray dose measurement by using the high-resolution characteristics of Gafchromic films in computed tomography. However, it is necessary to solve the problem of nonuniform thickness in the active layers of Gafchromic films. A double exposure technique using X-r...
متن کاملComparing three UV wavelengths for pre‐exposing Gafchromic EBT2 and EBT3 films
Gafchromic films are used for X-ray dose measurements during diagnostic examinations and have begun to be used for three-dimensional X-ray dose measurements using the high-resolution characteristics of Gafchromic films for computed tomography. However, the problem of unevenness in Gafchromic film active layers needs to be resolved. Double exposures using X-rays are performed during therapeutic ...
متن کاملCorrection of nonuniformity error of Gafchromic EBT2 and EBT3
This study investigates an X-ray dose measurement method for computed tomography using Gafchromic films. Nonuniformity of the active layer is a major problem in Gafchromic films. In radiotherapy, nonuniformity error is reduced by applying the double-exposure technique, but this is impractical in diagnostic radiology because of the heel effect. Therefore, we propose replacing the X-rays in the d...
متن کاملReflection spectrometry analysis of irradiated GAFCHROMIC XR type R radiochromic films.
The absorption spectra of the GAFCHROMIC XR type R radiochromic film measured with reflectance spectroscopy have been investigated in order to analyze the dosimetry characteristics of the film. Like the XR type T film, this film features two peaks in its absorption spectrum, approximately at 676 and 618 nm, whose intensities increase with increasing absorbed dose. When the main absorption peak ...
متن کاملبررسی میزان مواجهه شغلی با پرتوهای غیریونساز و اثربخشی وسایل حفاظت فردی در انواع جوشکاری قوس الکتریکی متداول
Introduction: Welding is considered to be one of the main sources of occupational exposure to non ionizing radiations. The present study was conducted to examine the rate of occupational exposure to non-ionizing radiation in common welding processes. Materials and Methods: In the present descriptive analytical study, occupational exposure to different types of non-ionizing radiations in...