Evaluation of the Effect of Tumor Position on Standardized Uptake Value Using Time-of-Flight Reconstruction and Point Spread Function

نویسندگان

  • Yasuharu Wakabayashi
  • Kenichi Kashikura
  • Yasuyuki Takahashi
  • Hitoshi Yabe
  • Akihiro Ichikawa
  • Souichi Yamamoto
  • Ayumi Ishii
  • Kunio Doi
چکیده

OBJECTIVES The present study was conducted to examine whether the standardized uptake value (SUV) may be affected by the spatial position of a lesion in the radial direction on positron emission tomography (PET) images, obtained via two methods based on time-of-flight (TOF) reconstruction and point spread function (PSF). METHODS A cylinder phantom with the sphere (30 mm diameter), located in the center was used in this study. Fluorine-18 fluorodeoxyglucose (18F-FDG) concentrations of 5.3 kBq/ml and 21.2 kBq/ml were used for the background in the cylinder phantom and the central sphere respectively. By the use of TOF and PSF, SUV max and SUV mean were determined while moving the phantom in a horizontal direction (X direction) from the center of field of view (FOV: 0 mm) at 50, 100, 150 and 200 mm positions, respectively. Furthermore, we examined 41 patients (23 male, 18 female, mean age: 68±11.2 years) with lymph node tumors, who had undergone 18F-FDG PET examinations. The distance of each lymph node from FOV center was measured, based on the clinical images. RESULTS As the distance of a lesion from the FOV center exceeded 100 mm, the value of SUV max , which was obtained with the cylinder phantom, was overestimated, while SUV mean by TOF and/or PSF was underestimated. Based on the clinical examinations, the average volume of interest was 8.5 cm3. Concomitant use of PSF increased SUV max and SUV mean by 27.9% and 2.8%, respectively. However, size of VOI and distance from the FOV center did not affect SUV max or SUV mean in clinical examinations. CONCLUSION The reliability of SUV quantification by TOF and/or PSF decreased, when the tumor was located at a 100 mm distance (or farther) from the center of FOV. In clinical examinations, if the lymph node was located within 100 mm distance from the center of FOV, SUV remained stable within a constantly increasing range by use of both TOF and PSF. We conclude that, use of both TOF and PSF may be helpful.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the Effect of Tumor Position on Standardized Uptake Value Using Time-of-Flight Reconstruction and Point Spread Function

Objective(s): The present study was conducted to examine whether the standardized uptake value (SUV) may be affected by the spatial position of a lesion in the radial direction on positron emission tomography (PET) images, obtained via two methods based on time-of-flight (TOF) reconstruction and point spread function (PSF). Methods: A cylinder phantom with the sphere (30mm diameter), located in...

متن کامل

Effect of Time-of-Flight and Regularized Reconstructions on Quantitative Measurements and Qualitative Assessments in Newly Diagnosed Prostate Cancer With 18F-Fluorocholine Dual Time Point PET/MRI

Recent technical advances in positron emission tomography/magnetic resonance imaging (PET/MRI) technology allow much improved time-of-flight (TOF) and regularized iterative PET reconstruction regularized iterative reconstruction (RIR) algorithms. We evaluated the effect of TOF and RIR on standardized uptake values (maximum and peak SUV [SUVmax and SUVpeak]) and their metabolic tumor volume depe...

متن کامل

Impact of Time-of-Flight and Point-Spread-Function for Respiratory Artifact Reduction in PET/CT Imaging: Focus on Standardized Uptake Value

Background The most important advantage of positron emission tomography/computed tomography (PET/CT) imaging is its capability of quantitative analysis. The aim of the current study was to choose the proper standardized uptake value (SUV) threshold, when the time-of-flight (TOF) and point spread function (PSF) were used for respiratory artifact reduction in the liver dome in a new-generation PE...

متن کامل

Edge Artifacts in Point Spread Function-based PET Reconstruction in Relation to Object Size and Reconstruction Parameters

Objective(s): We evaluated edge artifacts in relation to phantom diameter and reconstruction parameters in point spread function (PSF)-based positron emission tomography (PET) image reconstruction.Methods: PET data were acquired from an original cone-shaped phantom filled with 18F solution (21.9 kBq/mL) for 10 min using a Biograph mCT scanner. The images were reconstructed using the baseline or...

متن کامل

18F-FDG PET/CT of advanced gastric carcinoma and association of HER2 expression with standardized uptake value

Objective(s): Expression of HER2 in gastric carcinoma has direct prognostic and therapeutic implications in patient management. The aim of this study is to determine whether a relationship exists between standardized uptake value (SUV) and expression of HER2 in advanced gastric carcinoma. Methods: We analyzed the 18F-FDG PET/CT results of 109 patients that underwent gastrectomy for advanced gas...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 4  شماره 

صفحات  -

تاریخ انتشار 2016