Riata silicone defibrillation lead with normal electrical measures at routine ambulatory check: The role of high-voltage shock testing

نویسندگان

  • Elia De Maria
  • Ambra Borghi
  • Lorenzo Bonetti
  • Pier Luigi Fontana
  • Stefano Cappelli
چکیده

AIM To describe our experience with shock testing for the evaluation of patients with Riata™ leads. METHODS Among 51 patients with normal baseline electrical parameters, 20 died during follow-up. Of the remaining 31 patients, 15 underwent the test: In 10 cases a defibrillation testing with ventricular fibrillation (VF) induction and in 5 cases a R-wave-synchronized shock (> 20 J, without inducing VF). The test was performed under sedation with Midazolam. RESULTS Twelve patients (80%) had a normal behavior during shock testing: In 8 cases induced VF was correctly detected and treated; in 4 cases of R-wave-synchronized shock electrical parameters remained stable and normal. Three patients (20%) failed the test. One patient with externalized conductors showed a sudden drop of high-voltage impedance (< 10 Ohm) after a 25 J R-wave-synchronized shock. Two other patients with externalized conductors, undergoing defibrillation testing, showed a short-circuit during shock delivery and the implantable cardioverter defibrillator was unable to interrupt VF. CONCLUSION In Riata™ leads the delivery of a low current during routine measurement of high-voltage impedance may not reveal a small short circuit, that can only be evident by attempting to deliver a true shock, either for spontaneous arrhythmias or in the context of a shock testing.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Shock-induced right ventricular pacing failure caused by a short circuit: Uncommon but life-threatening complication of the Riata lead

Introduction The Riata and Riata ST implantable cardioverter-defibrillator (ICD) leads, a narrow caliber silicone lead without an outer jacket, have been shown to be prone to structural and electrical failure. While much interest has been shown in potentially lethal high-voltage (HV) failures caused by electrical short circuits due to a breach in the ethylene tetrafluoroethylene (ETFE) insulati...

متن کامل

Long-term performance of the St Jude Riata 1580-1582 ICD lead family.

OBJECTIVE Safety concerns about the Riata ICD shock lead were recently raised, with insulation failure due to conductor externalisation. Its incidence and presentation were assessed, and predictors of insulation failure and lead survival of the Riata 1580-1582 were studied, retrospectively, before the official recall. METHODS All 374 patients at the Erasmus Medical Center between July 2003 an...

متن کامل

Externalized Conductor Cables in QuickSite Left Ventricular Pacing Lead and Riata Right Ventricular Lead in a Single Patient: A Common Problem With Silicone Insulation

QuickSite (St Jude Medical, Sylmar, CA, USA) is a silicone and polyurethane-insulated coronary sinus pacing lead. Riata lead (St Jude Medical, Sylmar, CA, USA) is a silicone insulated right ventricular shock lead. Recently, insulation breach of silicone based leads raised a huge concern. Fluoroscopic examination of these two leads in the same patient revealed externalization of these two leads....

متن کامل

Fluoroscopic investigation of Riata® transvenous defibrillator leads.

BACKGROUND Implantable cardioverter-defibrillator leads from Riata® family (St. Jude Medical Inc., Sylmar, CA, USA) have been recently recalled by Food and Drug Administration for concerns of a unique type of "inside-out" insulation failure leading to conductor externalization. The objective of this study was to evaluate the prevalence and predictors of conductor externalization in patients imp...

متن کامل

Virtual electrode-induced phase singularity: a basic mechanism of defibrillation failure.

Delivery of a strong electric shock to the heart remains the only effective therapy against ventricular fibrillation. Despite significant improvements in implantable cardioverter defibrillator (ICD) therapy, the fundamental mechanisms of defibrillation remain poorly understood. We have recently demonstrated that a monophasic defibrillation shock produces a highly nonuniform epicardial polarizat...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 8  شماره 

صفحات  -

تاریخ انتشار 2016