A safe and effective method for pacemaker implantation in children Axillary vein puncture for device implantation and use of 4F catheter- delivered pacing leads in children

نویسندگان

  • Hania Burgan
  • Henri Sunthorn
  • Haran Burri
چکیده

Background: Contrast-guided axillary vein puncture has gained popularity for pacemaker and implantable cardioverter defibrillator (ICD) implantation in adults owing to its low risk profile, and could be an alternative to subclavian vein puncture, which is commonly performed in children. The Medtronic 3830 lead is a 4.1F lumenless catheter-delivered lead which may be particularly well-suited for use in the paediatric population because of the reduction in intravascular material and possibility for selective-site lead placement in complex anatomies. Methods: Data on paediatric patients at our institution aged <15 years who underwent transvenous device implantation using contrast-guided axillary vein puncture were retrieved. Contrast-guided axillary vein puncture was performed in all patients as the primary approach under general anaesthesia. Results: We retrieved data from 15 patients (7 males), aged 7.7 ± 4.1 yr (range 2–15 yr) at the time of intervention, and weighing 23.0 ± 9.6 kg (range 11.3–38.0 kg). Axillary vein puncture was successful in all patients. We placed 9 right atrial leads, 13 right ventricular pacing leads, 1 right ventricular ICD lead and 1 coronary sinus lead. A 3830 lead was successfully implanted in all 12 patients in whom this lead was intended. There were no procedure-related complications and lead electrical parameters were within normal limits in all patients after a mean follow-up of 13 months. Conclusion: Contrast-guided axillary vein puncture is a safe and effective method for pacemaker implantation in children. Furthermore, the thinbodied 3830 lead provides stable pacing parameters while reducing the bulk of intravascular material.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Axillary vein puncture without contrast venography for pacemaker and defibrillator leads implantation.

BACKGROUND Axillary vein puncture has been demonstrated to be an effective method for pacemaker and defibrillator leads implantation, without the complications encountered with the standard intrathoracic approach. OBJECTIVE Different techniques have been adopted for the cannulation of the axillary vein. We report our experience using the outer edge of the first rib below the inferior border o...

متن کامل

Left axillary pacemaker generator implantation with a direct puncture of the left axillary vein.

BACKGROUND Pacemaker generators are routinely implanted in the anterior chest. However, where to place the generator may need to be considered from the mental, functional, and cosmetic standpoints. METHODS In this study, we performed the left axillary pacemaker generator implantation with a direct puncture of the left axillary vein in 40 consecutive patients, and evaluated the late safety and...

متن کامل

Venous spasm during contrast-guided axillary vein puncture for pacemaker or defibrillator lead implantation.

AIMS Contrast-guided axillary vein puncture is commonly used for pacemaker or defibrillator lead implantation. Venous spasm during contrast-guided axillary vein puncture has been reported rarely. We investigated the incidence of venous spasm during contrast-guided axillary vein puncture and the impact of venous spasm on axillary vein puncture. METHODS AND RESULTS Seventy-four consecutive pati...

متن کامل

Comparison Between Open and Ultrasonography Guided Venous Access Ports in Children with Malignancy

Background: Long-term central venous access is used in children for various reasons specially for delivering chemotherapy. Since vessels in children have smaller diameters, they are more prone to injury and complications such as thrombosis. Different methods are used for implantation of port-a-cath in children. We aimed to compare the complications of insertion of central venous access ports be...

متن کامل

Pacing and defibrillation lead exchange without vein puncture.

BACKGROUND During lead implantation, venous access is generally achieved by puncturing the subclavian or axillary vein. Sometimes, although rarely, after lead positioning, the lead must be changed because of its inadequate mechanical stability or poor pacing parameters. This report concerns a technique of lead exchange that avoids an additional vein puncture. METHOD The tip of the lead, that ...