Evaluation of a new system, VITEK 2, for identification and antimicrobial susceptibility testing of enterococci.

نویسندگان

  • F Garcia-Garrote
  • E Cercenado
  • E Bouza
چکیده

We evaluated the new automated VITEK 2 system (bioMérieux) for the identification and antimicrobial susceptibility testing of enterococci. The results obtained with the VITEK 2 system were compared to those obtained by reference methods: standard identification by the scheme of Facklam and Sahm [R. R. Facklam and D. F. Sahm, p. 308-314, in P. R. Murray et al., ed., Manual of Clinical Microbiology, 6th ed., 1995] and with the API 20 STREP system and, for antimicrobial susceptibility testing, broth microdilution and agar dilution methods by the procedures of the National Committee for Clinical Laboratory Standards. The presence of vanA and vanB genes was determined by PCR. A total of 150 clinical isolates were studied, corresponding to 60 Enterococcus faecalis, 55 Enterococcus faecium, 26 Enterococcus gallinarum, 5 Enterococcus avium, 2 Enterococcus durans, and 2 Enterococcus raffinosus isolates. Among those isolates, 131 (87%) were correctly identified to the species level with the VITEK 2 system. Approximately half of the misidentifications were for E. faecium with low-level resistance to vancomycin, identified as E. gallinarum or E. casseliflavus; however, a motility test solved the discrepancies and increased the agreement to 94%. Among the strains studied, 66% were vancomycin resistant (57 VanA, 16 VanB, and 26 VanC strains), 23% were ampicillin resistant (MICs, >/=16 microgram/ml), 31% were high-level gentamicin resistant, and 45% were high-level streptomycin resistant. Percentages of agreement for susceptibility and resistance to ampicillin, vancomycin, and teicoplanin and for high-level gentamicin resistance and high-level streptomycin resistance were 93, 95, 97, 97, and 96%, respectively. The accuracy of identification and antimicrobial susceptibility testing of enterococci with the VITEK 2 system, together with the significant reduction in handling time, will have a positive impact on the work flow of the clinical microbiology laboratory.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Validation of VITEK 2 version 4.01 software for detection, identification, and classification of glycopeptide-resistant enterococci.

We evaluated the ability of the new VITEK 2 version 4.01 software to identify and detect glycopeptide-resistant enterococci compared to that of the reference broth microdilution method and to classify them into the vanA, vanB, vanC1, and vanC2 genotypes. Moreover, the accuracy of antimicrobial susceptibility testing with agents with improved potencies against glycopeptide-resistant enterococci ...

متن کامل

Performance of Vitek 2 for antimicrobial susceptibility testing of Staphylococcus spp. and Enterococcus spp.

Vitek 2 (bioMérieux, Inc., Durham, NC) is a widely used commercial antimicrobial susceptibility testing system. We compared MIC results obtained by Vitek 2 to those obtained by the Clinical and Laboratory Standards Institute (CLSI) broth microdilution (BMD) reference method for 134 staphylococcal and 84 enterococcal clinical isolates. Nineteen agents were evaluated, including all those availabl...

متن کامل

Accuracy of the VITEK 2 system to detect glycopeptide resistance in enterococci.

We evaluated the accuracy of the VITEK 2 fully automated system to detect and identify glycopeptide-resistant enterococci (GRE) compared to a reference agar dilution method. The sensitivity of vancomycin susceptibility testing with VITEK 2 for the detection of vanA, vanB, and vanC1 strains was 100%. The sensitivity of vancomycin susceptibility testing of vanC2 strains was 77%. The sensitivity o...

متن کامل

Evaluation of Vitek system for susceptibility testing of Enterococcus faecium isolates.

Although imipenem is not a first-line drug for treating enterococcal infection, it could well become a useful drug for treating mixed infections that include enterococci. However, there is no NCCLS guideline for susceptibility testing of imipenem versus enterococci. Moreover, there are no statements to indicate that in vitro susceptibility results for other antimicrobial agents can be used to p...

متن کامل

Comparative evaluation of the VITEK 2 Advanced Expert System (AES) in five UK hospitals.

OBJECTIVES We carried out an evaluation of VITEK 2 in five UK laboratories, comparing results with 'gold standard' agar-dilution MIC data, assessing its ability to recognize resistant phenotypes and comparing results with those generated by routine antimicrobial susceptibility testing methods. METHODS Laboratories tested a collection of 82 strains selected on the basis of their challenging an...

متن کامل