Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.
نویسندگان
چکیده
BACKGROUND Gross trunnion failure (GTF) is a rare complication in total hip arthroplasty (THA) reported across a range of manufacturers. Specific lots of the Stryker low friction ion treatment (LFIT) anatomic cobalt chromium alloy (CoCr) V40 femoral head were recalled in August 2016. In part, the recall was based out of concerns for disassociation of the femoral head from the stem and GTF. METHODS We report on 28 patients (30 implants) with either GTF (n = 18) or head-neck taper corrosion (n = 12) of the LFIT CoCr femoral head and the Accolade titanium-molybdenum-zirconium-iron alloy femoral stems. All these cases were associated with adverse local tissue reactions requiring revision of the THA. RESULTS In our series, a conservative estimate of the incidence of failure was 4.7% (n = 636 total implanted) at 8.0 ± 1.4 years from the index procedure. Failures were associated with a high-offset 127° femoral stem neck angle and increased neck lengths; 43.3% (13 of 30) of the observed failures included implant sizes outside the voluntary recall (27.8% [5 of 18] of the GTF and 75.0% [8 of 12] of the taper corrosion cases). Serum cobalt and chromium levels were elevated (cobalt: 8.4 ± 7.0 μg/mL; chromium: 3.4 ± 3.3 μ/L; cobalt/chromium ratio: 3.7). The metal artifact reduction sequence magnetic resonance imaging demonstrated large cystic fluid collections typical with adverse local tissue reactions. During revision, a pseudotumor was observed in all cases. Pathology suggested a chronic inflammatory response. Impending GTF could be diagnosed based on aspiration of black synovial fluid and an oblique femoral head as compared with the neck taper on radiographs. CONCLUSION In our series of the recalled LFIT CoCr femoral head, the risk of impending GTF or head-neck taper corrosion should be considered as a potential diagnosis in a painful LFIT femoral head and Accolade titanium-molybdenum-zirconium-iron alloy THA with unknown etiology. Almost half of the failures we observed included sizes outside of the voluntary recall.
منابع مشابه
Catastrophic Femoral Head-Stem Trunnion Dissociation Secondary to Corrosion.
BACKGROUND Modular femoral heads provide procedural enhancement by allowing accurate restoration of hip offset and limb-length equalization. However, corrosion may lead to adverse local tissue reactions. Severe trunnion corrosion can also lead to femoral head dissociation and catastrophic implant failure following primary total hip arthroplasty. METHODS We describe 5 cases, from our instituti...
متن کاملCorrosion and adverse tissue reaction after modular unipolar hip hemiarthroplasty
Modern unipolar hip hemiarthroplasty, commonly used for displaced femoral neck fracture, is now modular, with both a variable length cobalt-chromium adapter-sleeve and large femoral head. Patients with these modular components may develop symptomatic trunnion corrosion, with elevated serum metal levels. We report the case of an 82-year-old woman, 5.5 years after a modular unipolar hip hemiarthr...
متن کاملEffect of carbon ion implantation on the tribology of metal-on-metal bearings for artificial joints
Metal-on-metal (MoM) bearings have become popular due to a major advantage over metal-on-polymer bearings for total hip arthroplasty in that the larger femoral head and hydrodynamic lubrication of the former reduce the rate of wear. However, concerns remain regarding adverse reactions to metal debris including metallosis caused by metal wear generated at the taper-head interface and another mod...
متن کاملEvora® chromium-cobalt dual mobility socket: results at a minimum 10 years' follow-up.
INTRODUCTION The Evora chromium-cobalt alloy dual mobility socket claims to display a large articulation tribology different from that of stainless steel models, limiting the risk of intraprosthetic dislocation and wear. The present study reports a minimum of 10years' follow-up in a multicenter prospective series of 200 sockets previously reported on at 5years. HYPOTHESIS The use of chromium-...
متن کاملLarge head metal-on-metal cementless total hip arthroplasty versus 28mm metal-on-polyethylene cementless total hip arthroplasty: design of a randomized controlled trial
BACKGROUND Osteoarthritis of the hip is successfully treated by total hip arthroplasty with metal-on-polyethylene articulation. Polyethylene wear debris can however lead to osteolysis, aseptic loosening and failure of the implant. Large head metal-on-metal total hip arthroplasty may overcome polyethylene wear induced prosthetic failure, but can increase systemic cobalt and chromium ion concentr...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of arthroplasty
دوره 32 9 شماره
صفحات -
تاریخ انتشار 2017