Clinical evaluation of a new in-line continuous blood gas monitor.

نویسندگان

  • R Southworth
  • R Sutton
  • S Mize
  • A H Stammers
  • L W Fristoe
  • D Cook
  • D Hostetler
  • W E Richenbacher
چکیده

Two methodologies for obtaining accurate blood gas and electrolyte values during cardiopulmonary bypass (CPB) are traditional laboratory analyzers, which use an electrochemical technology, and continuous in-line monitoring systems, which use a fluorometric and/or spectrophotometric technology. The purpose of the present study was to evaluate the accuracy of a new continuous in-line monitor, the 3M CDI Blood Parameter Monitoring System 500, which provides continuous in-line measurements of pH, PCO2, PO2, potassium (K+), oxygen saturation, hematocrit, hemoglobin, and temperature, during partial or complete CPB. Study parameters included arterial pH, PCO2, PO2, and K+ values. Overall performance was analyzed by calculating the mean difference (expressed as the bias) between the CDI system 500 and the laboratory analyzer for each parameter. The accuracy of the arterial pH, PCO2, and K+ values provided by the CDI system 500 was then evaluated using target values established in the acceptable performance standards for laboratory analyzers from the Clinical Laboratory Improvement Act of 1988 (CLIA '88). The accuracy of the PO2 value provided by the CDI system 500 was evaluated using a target value of +/- 10% of the reference, or laboratory analyzer, value. A prospective multi-center trial was conducted following Institutional Review Board approval. A total of 75 cases was included in the analyses, with over 200 data points from 4 clinical locations. Results for pH, PCO2, and K+ were within the target values established by CLIA '88. pH bias was 0.00 +/- 0.02 pH units. PCO2 bias was -0.3 +/- 3.3 mm Hg. K+ bias was approximately +0.12 +/- 0.31 mmole/l. Results for PO2 were within 10% of the reference value. PO2 bias was 7.5 +/- 13.8 mm Hg. The results of this clinical trial show that the CDI System 500 continuous in-line monitoring system provides values that meet the accuracy standards for laboratory analyzers for arterial pH, PCO2, PO2, and K+.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Anatomical parameters of cardiopulmonary system, venous blood pCO2and pO2tensions and the development of ascites syndrome in two genetic line chickens that are differing in their growth rate

The present study was designed to examine the anatomical parameters of the cardiopulmonary system, the function of venous blood gas parameters and the development of ascites incidence in two genetic line chickens. Three hundred forty day-old chickens from two pure broiler breeder lines, which were different in their growth rate and susceptibility to ascites syndrome were obtained. The relative ...

متن کامل

Clinical evaluation of the CDI-100 in-line hematocrit/saturation monitor.

This study was undertaken to evaluate the accuracy, reliability, consistency and biases of the CDI-100 saturation monitor when compared with a blood gas analyzer. The advantage of continuous in-line monitoring is that the perfusionist has continuous updates as to the patient's changing physiologic state. During this study, if the sample readout of the CDI-100 was off by greater than 10% from th...

متن کامل

Clinical evaluation of the accuracy and precision of the CDI 500 in-line blood gas monitor with and without gas calibration.

During cardiopulmonary bypass blood gases can be analyzed with laboratory equipment or with an in-line monitor giving instant results. The manufacturer of the CDI 500 in-line blood gas monitor recommends gas calibration before use. In acute cases there may not be time to perform a gas calibration. We hypothesized that after calibration against laboratory results, the CDI values of pH, pO2, and ...

متن کامل

The effects of continuous blood gas monitoring during cardiopulmonary bypass: a prospective, randomized study--Part I.

The use of continuous in-line blood gas management (CILBGM) is steeped in controversy concerning its potential utility and impact on patient outcomes. The purpose of this study was to determine whether the use of CILBGM results in improved quality of patient care. Fifty-nine patients were enrolled in a Institutional Review Board-approved, prospective, randomized study. An in-line blood gas moni...

متن کامل

مقایسه اندازه‌گیری اشباع اکسیژن خون شریانی با پالس اکسی‌متری و نمونه خون شریانی در بیماران بخش مراقبت‌های ویژه قلبی

Background: Pulseoximetry is widely used in the critical care setting, currently used to guide therapeutic interventions. Few studies have evaluated the accuracy of SPO2 (puls-eoximetry oxygen saturation) in intensive care unit after cardiac surgery. Our objective was to compare pulseoximetry with arterial oxygen saturation (SaO2) during clinical routine in such patients, and to examine the eff...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The journal of extra-corporeal technology

دوره 30 4  شماره 

صفحات  -

تاریخ انتشار 1998