Assessing respiratory function depends on mechanical characteristics of balloon catheters.
نویسندگان
چکیده
BACKGROUND Respiratory muscle function and lung and chest wall mechanics are reliably assessed by esophageal and gastric balloon catheters. The aim of this in vitro bench study was to assess the mechanical properties of commercially available balloon catheters using an experimental model with 3 defined compliances (27, 54, 90 mL/cm H2O). METHODS Six catheters were investigated in 4 conditions: (1) balloon pressure during initial inflation, (2) static pressure measurements at different filling volumes, (3) estimation of set compliances in the experimental lung model at different levels of superimposed pressure, and (4) elastic balloon properties after 16 h of inflation. RESULTS 5/6 catheters showed initial pressure artifacts resulting from material adhesion. All static pressure measurements could be performed with an error < 1 cm H2O. Balloon overfilling resulted in larger errors in 4/6 catheters. Compliance determined from pressure measurements via the catheters differed by < 5% from that determined from direct pressure measurements. Sixteen hours of inflation resulted in a broader working range, that is, overfilling effects occurred at higher filling volumes. CONCLUSIONS The reliability of pressure measurements and estimation of the lung model's compliance in the tested catheters are high. Filling volume appears to be critical for precise pressure measurement and compliance estimation. At first use, adhesion of the balloon material might prevent reliable pressure measurement.
منابع مشابه
Response characteristics of esophageal balloon catheters handmade using latex and nonlatex materials
The measurement of esophageal pressure allows for the calculation of several important and clinically useful parameters of respiratory mechanics. Esophageal pressure is often measured with balloon-tipped catheters. These catheters may be handmade from natural latex condoms and polyethylene tubing. Given the potential of natural latex to cause allergic reaction, it is important to determine whet...
متن کاملEsophageal and gastric pressure measurements.
The measurement of esophageal and gastric pressures with balloon-tipped catheters has been used with great success over the past half century to delineate the physiology of the mechanical respiratory system. Pleural pressure and abdominal pressure values estimated from esophageal and gastric pressure measurements allow analysis of lung and chest wall compliance, as well as work of breathing, re...
متن کاملAn in vivo comparison of a catheter mounted pressure transducer system with conventional balloon catheters.
In the assessment of respiratory muscle function balloon catheters have been widely used for pressure measurements. However, this type of investigation is poorly tolerated by acutely ill patients. This study assessed the performance of a possible alternative, a catheter-mounted miniature pressure transducer (CMT). The assessment consisted of a laboratory study of the linearity, frequency respon...
متن کاملLiquid- and Air-Filled Catheters without Balloon as an Alternative to the Air-Filled Balloon Catheter for Measurement of Esophageal Pressure
BACKGROUND Measuring esophageal pressure (Pes) using an air-filled balloon catheter (BC) is the common approach to estimate pleural pressure and related parameters. However, Pes is not routinely measured in mechanically ventilated patients, partly due to technical and practical limitations and difficulties. This study aimed at comparing the conventional BC with two alternative methods for Pes m...
متن کاملDevelopment of a balloon volume sensor for pulsating balloon catheters.
Helium pulsed balloons are integral components of several cardiovascular devices, including intraaortic balloon pumps (IABP) and a novel intravenous respiratory support catheter. Effective use of these devices clinically requires full inflation and deflation of the balloon, and improper operating conditions that lead to balloon under-inflation can potentially reduce respiratory or cardiac suppo...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Respiratory care
دوره 59 9 شماره
صفحات -
تاریخ انتشار 2014