The effect of steam sterilization on the accuracy of spring-style mechanical torque devices for dental implants
نویسندگان
چکیده
BACKGROUND Mechanical torque devices (MTDs) are one of the most commonly recommended devices used to deliver optimal torque to the screw of dental implants. Recently, high variability has been reported about the accuracy of spring-style mechanical torque devices (S-S MTDs). Joint stability and survival rate of fixed implant supported prosthesis depends on the accuracy of these devices. Currently, there is limited information on the steam sterilization influence on the accuracy of MTDs. The purpose of this study was to assess the effect of steam sterilization on the accuracy (±10% of the target torque) of spring-style mechanical torque devices for dental implants. MATERIALS AND METHODS Fifteen new S-S MTDs and their appropriate drivers from three different manufacturers (Nobel Biocare, Straumann [ITI], and Biomet 3i [3i]) were selected. Peak torque of devices (5 in each subgroup) was measured before and after autoclaving using a Tohnichi torque gauge. Descriptive statistical analysis was used and a repeated-measures ANOVA with type of device as a between-subject comparison was performed to assess the difference in accuracy among the three groups of spring-style mechanical torque devices after sterilization. A Bonferroni post hoc test was used to assess pairwise comparisons. RESULTS Before steam sterilization, all the tested devices stayed within 10% of their target values. After 100 sterilization cycles, results didn't show any significant difference between raw and absolute error values in the Nobel Biocare and ITI devices; however the results demonstrated an increase of error values in the 3i group (P < 0.05). Raw error values increased with a predictable pattern in 3i devices and showed more than a 10% difference from target torque values (maximum difference of 14% from target torque was seen in 17% of peak torque measurements). CONCLUSION Within the limitation of this study, steam sterilization did not affect the accuracy (±10% of the target torque) of the Nobel Biocare and ITI MTDs. Raw error values increased with a predictable pattern in 3i devices and showed more than 10% difference from target torque values. Before expanding upon the clinical implications, the controlled and combined effect of aging (frequency of use) and steam sterilization needs more investigation.
منابع مشابه
The combined effect of dismantling for steam sterilization and aging on the accuracy of spring-style mechanical torque devices
PURPOSE This study aimed to assess the combined effect of dismantling before sterilization and aging on the accuracy (±10% of the target torque) of spring-style mechanical torque devices (S-S MTDs). METHODS Twenty new S-SMTDs from two different manufacturers (Nobel Biocare and Straumann: 10 of each type) were selected and divided into two groups, namely, case (group A) and control (group B). ...
متن کاملEvaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants
OBJECTIVE Accurate delivery of torque to implant screws is critical to generate ideal preload in the screw joint and to offer protection against screw loosening. Mechanical torque-limiting devices (MTLDs) are available for this reason. In this study, the accuracy of one type of friction-style and two types of spring-style MTLDs at baseline, following fatigue conditions and sterilization process...
متن کاملThe effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
BACKGROUND Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devic...
متن کاملEffect of Component Dismantling on Quality of Sterilization of Mechanical Torque Limiting Devices(Pilot Study)
Objectives Dental instruments may serve as a route of infection transmission. The necessity of dismantling of Friction-Style (FS) and SpringStyle (SS) mechanical torque limiting devices (MTLDs) prior to disinfection and sterilization is a matter of debate, aside fromthe fact that it is difcult and time-consuming. This study sought to assess the effect of dismantling of components on quality of ...
متن کاملبررسی تأثیر دفعات کارکرد بر دقت محدودکنندههای مکانیکی تورک از نوع فنری در ایمپلنتهای دندانی□
Objective: Recently, a wide variation has been reported in delivery of the target torque by spring-style mechanical torque limiting devices (MTLDs) but the effect of aging on their accuracy has not been independently evaluated. The purpose of this study was to determine the effect of aging on the accuracy of spring-style MTLDs. Methods: In this in-vitro experimental study, 15 new spring-style M...