Measurement Accuracy in 3D Electromagnetic Articulography (Carstens AG500)

نویسنده

  • Christian Kroos
چکیده

In this study the measurement accuracy of the 3D EMA (Electromagnetic Articulograhy) system Carstens AG500 was evaluated using an optical motion capture system (Vicon). A custom-built container fixed the positions of the EMA sensors and the optical markers and a series of movements of the container within the EMA measurement field was recorded. The results show that (a) the very small error of static position measurements does not reflect the behaviour of the system in dynamic measurements and (b) that in dynamic measurements the magnitude of the error varies substantially across the three spatial dimensions.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Electromagnetic articulography with AG500 and AG501

The AG500 (Carstens Medizinelektronik GmbH) is widely used in electromagnetic articulography, as it allows the recording in 3D of the movements of the articulatory organs. However some anomalies in sensor position tracking were recognized and analyzed in previous works, which ultimately attributed them to numerical issues and not to external interferences. Recently a new upgraded model, the AG5...

متن کامل

Learning the Mapping Function from Voltage Amplitudes to Sensor Positions in 3D-EMA Using Deep Neural Networks

The first generation of three-dimensional Electromagnetic Articulography devices (Carstens AG500) suffered from occasional critical tracking failures. Although now superseded by new devices, the AG500 is still in use in many speech labs and many valuable data sets exist. In this study we investigate whether deep neural networks (DNNs) can learn the mapping function from raw voltage amplitudes t...

متن کامل

Announcing the Electromagnetic Articulography (Day 1) Subset of the family mngu0 Articulatory Corpus

This paper serves as an initial announcement of the availability of a corpus of articulatory data called mngu0. This corpus will ultimately consist of a collection of multiple sources of articulatory data acquired from a single speaker: electromagnetic articulography (EMA), audio, video, volumetric MRI scans, and 3D scans of dental impressions. This data will be provided free for research use. ...

متن کامل

Aspects of pharyngealized phonemes in Arabic using articulography

In this paper, we present preliminary results of an articulatory study of pharyngealized phonemes in Arabic using an electromagnetic articulograph (AG500). We compared the articulations of pharyngealized phonemes to non-pharyngealized ones. The articulation of the tongue was tracked by four sensors glued onto the tongue. A corpus of several CVCVCVs was recorded using the AG500, labeled, and ana...

متن کامل

Accuracy assessment for AG500, electromagnetic articulograph.

PURPOSE The goal of this article was to evaluate the accuracy and reliability of the AG500 (Carstens Medizinelectronik, Lenglern, Germany), an electromagnetic device developed recently to register articulatory movements in three dimensions. This technology seems to have unprecedented capabilities to provide rich information about time-varying positions of articulators. However, strengths and we...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2009