Electron contamination modeling and skin dose in 6 MV longitudinal field MRIgRT: Impact of the MRI and MRI fringe field.
نویسندگان
چکیده
PURPOSE In recent times, longitudinal field MRI-linac systems have been proposed for 6 MV MRI-guided radiotherapy (MRIgRT). The magnetic field is parallel with the beam axis and so will alter the transport properties of any electron contamination particles. The purpose of this work is to provide a first investigation into the potential effects of the MR and fringe magnetic fields on the electron contamination as it is transported toward a phantom, in turn, providing an estimate of the expected patient skin dose changes in such a modality. METHODS Geant4 Monte Carlo simulations of a water phantom exposed to a 6 MV x-ray beam were performed. Longitudinal magnetic fields of strengths between 0 and 3 T were applied to a 30 × 30 × 20 cm(3) phantom. Surrounding the phantom there is a region where the magnetic field is at full MRI strength, consistent with clinical MRI systems. Beyond this the fringe magnetic field entering the collimation system is also modeled. The MRI-coil thickness, fringe field properties, and isocentric distance are varied and investigated. Beam field sizes of 5 × 5, 10 × 10, 15 × 15 and 20 × 20 cm(2) were simulated. Central axis dose, 2D virtual entry skin dose films, and 70 μm skin depth doses were calculated using high resolution scoring voxels. RESULTS In the presence of a longitudinal magnetic field, electron contamination from the linear accelerator is encouraged to travel almost directly toward the patient surface with minimal lateral spread. This results in a concentration of electron contamination within the x-ray beam outline. This concentration is particularly encouraged if the fringe field encompasses the collimation system. Skin dose increases of up to 1000% were observed for certain configurations and increases above Dmax were common. In nonmagnetically shielded cases, electron contamination generated from the jaw faces and air column is trapped and propagated almost directly to the phantom entry region, giving rise to intense dose hot spots inside the x-ray treatment field. These range up to 1000% or more of Dmax at the CAX, depending on field size, isocenter, and coil thickness. In the case of a fully magnetically shielded collimation system and the lowest MRI field of 0.25 T, the entry skin dose is expected to increase to at least 40%, 50%, 65%, and 80% of Dmax for 5 × 5, 10 × 10, 15 × 15, and 20 × 20 cm(2), respectively. CONCLUSIONS Electron contamination from the linac head and air column may cause considerable skin dose increases or hot spots at the beam central axis on the entry side of a phantom or patient in longitudinal field 6 MV MRIgRT. This depends heavily on the properties of the magnetic fringe field entering the linac beam collimation system. The skin dose increase is also related to the MRI-coil thickness, the fringe field, and the isocenter distance of the linac. The results of this work indicate that the properties of the MRI fringe field, electron contamination production, and transport must be considered carefully during the design stage of a longitudinal MRI-linac system.
منابع مشابه
Estimation of Secondary Skin Cancer Risk Due To Electron Contamination in 18-MV LINAC-Based Prostate Radiotherapy
Introduction Accurate estimation of the skin-absorbed dose in external radiation therapy is essential to estimating the probability of secondary carcinogenesis induction Materials and Methods Electron contamination in prostate radiotherapy was investigated using the Monte Carlo (MC) code calculation. In addition, field size dependence of the skin dose was assessed. Excess cancer risk induced by...
متن کاملMonte Carlo estimation of electron contamination in a 18 MV clinical photon beam
Background: The electron contamination may reduce or even diminish the skin sparing property of the megavoltage beam. The detailed characteristics of contaminant electrons are presented for different field sizes and cases. Materials and Methods: The Monte Carlo code, MCNPX, has been used to simulate 18 MV photon beam from a Varian Linac-2300 accelerator. All dose measurements were car...
متن کاملEvaluation of MRI-based MAGIC polymer gel dosimeter in small photon fields
Background: Accurate small radiation field dosimetry is essential in modern radiotherapy techniques such as stereotactic radiosurgery (SRS) and intensity modulated radiotherapy (IMRT). Precise measurement of dosimetric parameters such as beam profile, percentage depth doses and output factor of these beams are complicated due to the electron disequilibrium and the steep dose gradients. In the p...
متن کاملSkin dose in longitudinal and transverse linac-MRIs using Monte Carlo and realistic 3D MRI field models.
PURPOSE The magnetic fields of linac-MR systems modify the path of contaminant electrons in photon beams, which alters patient skin dose. To accurately quantify the magnitude of changes in skin dose, the authors use Monte Carlo calculations that incorporate realistic 3D magnetic field models of longitudinal and transverse linac-MR systems. METHODS Finite element method (FEM) is used to genera...
متن کاملCalculation of total dose and dose equivalent distribution in the treatment of lung cancer using MR-guided carbon therapy
Nowadays, in order to improve the accuracy of treatment in radiation therapy, there are many attempts to use magnetic resonance imaging (MRI) due to the advantages of excellent soft tissue contrast and ultra-fast pulse sequences. On the other hand, carbon-ion radiation therapy is developing rapidly due to the benefits of greater relative biological effectiveness (RBE) and the application in the...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 39 2 شماره
صفحات -
تاریخ انتشار 2012