Projection-based estimation and nonuniformity correction of sensitivity profiles in phased-array surface coils.

نویسندگان

  • Sungdae Yun
  • Walid E Kyriakos
  • Jun-Young Chung
  • Yeji Han
  • Seung-Schik Yoo
  • Hyunwook Park
چکیده

PURPOSE To develop a novel approach for calculating the accurate sensitivity profiles of phased-array coils, resulting in correction of nonuniform intensity in parallel MRI. MATERIALS AND METHODS The proposed intensity-correction method estimates the accurate sensitivity profile of each channel of the phased-array coil. The sensitivity profile is estimated by fitting a nonlinear curve to every projection view through the imaged object. The nonlinear curve-fitting efficiently obtains the low-frequency sensitivity profile by eliminating the high-frequency image contents. Filtered back-projection (FBP) is then used to compute the estimates of the sensitivity profile of each channel. The method was applied to both phantom and brain images acquired from the phased-array coil. RESULTS Intensity-corrected images from the proposed method had more uniform intensity than those obtained by the commonly used sum-of-squares (SOS) approach. With the use of the proposed correction method, the intensity variation was reduced to 6.1% from 13.1% of the SOS. When the proposed approach was applied to the computation of the sensitivity maps during sensitivity encoding (SENSE) reconstruction, it outperformed the SOS approach in terms of the reconstructed image uniformity. CONCLUSION The proposed method is more effective at correcting the intensity nonuniformity of phased-array surface-coil images than the conventional SOS method. In addition, the method was shown to be resilient to noise and was successfully applied for image reconstruction in parallel imaging.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Improvement of brain segmentation accuracy by optimizing non-uniformity correction using N3

Smoothly varying and multiplicative intensity variations within MR images that are artifactual, can reduce the accuracy of automated brain segmentation. Fortunately, these can be corrected. Among existing correction approaches, the nonparametric non-uniformity intensity normalization method N3 (Sled, J.G., Zijdenbos, A.P., Evans, A.C., 1998. Nonparametric method for automatic correction of inte...

متن کامل

High-sensitivity coil array for head and neck imaging: technical note.

The purpose of this study was to develop coils for MR imaging of the head and neck region, with the aim of improving sensitivity and coverage. A head and neck phased array coil was constructed and compared with volume and temporomandibular joint surface coils for sensitivity and coverage in phantom studies. An algorithm was implemented to correct for the nonuniformity in the surface coil recept...

متن کامل

Cardiovascular M R I probes for the outs ide in and for the inside

Proton (1H) magnetic resonance imaging (MRI), is now an established clinical diagnostic tool, and 25 years of steady progress have reduced MRI scan-times of an hour or more to sub-seconds: MRI fluoroscopy is feasible! Arguably, advances in MRI detector coil design are major factors in the tremendous gains in sensitivity that have made this possible. MRI detectors comprised of small surface coil...

متن کامل

Intensity Correction in Intravascular MRI using Projection Images

Introduction: Potential clinical benefits of high-resolution imaging using intravascular probes is often limited due to the image intensity distortion caused by the signal inhomogeneity associated with the placement of the probe. Methods developed for the phased-array and other surface coils are not as successful for intravascular imaging since the conducting structures of the RF-receiving devi...

متن کامل

In vivo method for correcting transmit/receive nonuniformities with phased array coils.

Phased array coils are finding widespread applications in both the research and the clinical setting. However, intensity nonuniformities with such coils can reduce the potential benefits of these coils, particularly for applications such as tissue segmentation. In this work, a method is described for correcting the nonuniform signal response based on in vivo measures of both the transmission fi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of magnetic resonance imaging : JMRI

دوره 25 3  شماره 

صفحات  -

تاریخ انتشار 2007