Detection of Atrial Fibrillation and Flutter by a Dual-Chamber Implantable Cardioverter-Defibrillator
نویسندگان
چکیده
Background—To distinguish prolonged episodes of atrial fibrillation (AF) that require cardioversion from self-terminating episodes that do not, an atrial implantable cardioverter-defibrillator (ICD) must be able to detect AF continuously for extended periods. The ICD should discriminate between atrial tachycardia/flutter (AT), which may be terminated by antitachycardia pacing, and AF, which requires cardioversion. Methods and Results—We studied 80 patients with AT/AF and ventricular arrhythmias who were treated with a new atrial/dual-chamber ICD. During a follow-up period lasting 662 months, we validated spontaneous, device-defined AT/AF episodes by stored electrograms in all patients. In 58 patients, we performed 80 Holter recordings with telemetered atrial electrograms, both to validate the continuous detection of AT/AF and to determine the sensitivity of the detection of AT/AF. Detection was appropriate in 98% of 132 AF episodes and 88% of 190 AT episodes (98% of 128 AT episodes with an atrial cycle length ,300 ms). Intermittent sensing of far-field R waves during sinus tachycardia caused 27 inappropriate AT/AF detections; these detections lasted 2.662.0 minutes. AT/AF was detected continuously in 27 of 28 patients who had spontaneous episodes of AT/AF (96%). The device memory recorded 90 appropriate AT/AF episodes lasting .1 hour, for a total of 2697 hours of continuous detection of AT/AF. During Holter monitoring, the sensitivity of the detection of AT/AF (116 hours) was 100%; the specificity of the detection of non-AT/AF rhythms (1290 hours) was 99.99%. Of 166 appropriate episodes detected as AT, 45% were terminated by antitachycardia pacing. Conclusions—A new ICD detects AT/AF accurately and continuously. Therapy may be programmed for long-duration AT/AF, with a low risk of underdetection. Discrimination of AT from AF permits successful pacing therapy for a significant fraction of AT. (Circulation. 2000;101:878-885.)
منابع مشابه
Implantable cardioverter-defibrillator in a patient with dextrocardia situs inversus
Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...
متن کاملInappropriate cardioverter-defibrillator discharge continues to be a major problem in clinical practice.
BACKGROUND The purpose of this study was to determine the rate and causes of inappropriate rhythm detection, and to compare adequacy of ventricular arrhythmia detection by single-chamber and dual-chamber cardioverter-defibrillators (ICD). METHODS We evaluated 190 patients (age 57.2 +/- 11.2 years) with ICD. FOLLOW-UP 34.3 +/- +/- 22 months. Dual-chamber ICD was used in 54 patients. RESULT...
متن کاملPrevention of inappropriate therapy in implantable cardioverter-defibrillators: results of a prospective, randomized study of tachyarrhythmia detection algorithms.
OBJECTIVES The purpose of this randomized study was to investigate the performance of single- and dual-chamber tachyarrhythmia detection algorithms. BACKGROUND A proposed benefit of dual-chamber implantable cardioverter-defibrillators (ICDs) is improved specificity of tachyarrhythmia detection. METHODS All ICD candidates received a dual-chamber ICD and were randomized to programmed single- ...
متن کاملAtrial Sensing and Blanking in Dual-Chamber Implantable Cardioverter-Defibrillators
Atrial Sensing and Blanking in Dual-Chamber Implantable Cardioverter-Defibrillators Accurate atrial sensing during tachycardia is essential for reliable determination of atrial rate, which is a critical component of interval-based dual-chamber supraventricular tachycardia (SVT)–VT discrimination algorithms. Thus, atrial sensing in dual-chamber implantable cardioverter-defibrillators (ICDs) must...
متن کاملDetection of atrial fibrillation and flutter by a dual-chamber implantable cardioverter-defibrillator. For the Worldwide Jewel AF Investigators.
BACKGROUND To distinguish prolonged episodes of atrial fibrillation (AF) that require cardioversion from self-terminating episodes that do not, an atrial implantable cardioverter-defibrillator (ICD) must be able to detect AF continuously for extended periods. The ICD should discriminate between atrial tachycardia/flutter (AT), which may be terminated by antitachycardia pacing, and AF, which req...
متن کامل