BACKGROUND Catheter ablation of atrial fibrillation is currently guided by x-ray fluoroscopy. The associated radiation risk to patients and medical staff may be significant. We report an atrial fibrillation ablation technique using intracardiac echocardiography (ICE) and electroanatomic mapping without fluoroscopy. METHODS AND RESULTS Twenty-one patients with atrial fibrillation (age, 42 to 73 years; 14 male; 14 paroxysmal, 7 persistent; body mass index, 26 to 38) underwent ablation. A decapolar catheter was advanced through the left subclavian vein until stable coronary sinus electrograms appeared on all electrodes. Two 9F sheaths were advanced transfemorally over a guide wire to the right atrium. A rotational ICE catheter was advanced through a deflectable sheath. Double transseptal puncture was performed with ICE guidance and facilitated by electrocautery. A 3D MRI left atrial image was registered to the ostia of the pulmonary veins using ICE. Catheter ablation was performed using ICE and electroanatomic mapping navigation. In 19 cases, no fluoroscopy was used and the staff did not wear protective lead. In 2 cases, 2 to 16 minutes of fluoroscopy was used to assist transseptal puncture. Median procedure time was 208 (188 to 221) minutes; coronary sinus cannulation took 5 (2 to 26) minutes; double transseptal took 26 (17 to 40) minutes; left atrial ablation time was 103 (90 to 127) minutes. All patients underwent circumferential pulmonary vein ablation and 8 patients underwent additional left atrial ablation. There were no procedure-related complications. CONCLUSIONS Catheter ablation of atrial fibrillation without fluoroscopy is feasible and merits further attention. This technique may be especially helpful in preventing x-ray exposure in children, pregnant women, and obese patients undergoing left atrial ablation.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Intracardiac ECHO Integration with Three Dimensional Mapping: Role in AF Ablation.

Background: Catheter ablation of atrial fibrillation (AF) is typically guided by 3D mapping. This involves point-by-point reconstruction of the 3D virtual anatomy and may be time consuming and require substantial fluoroscopy exposure. Intracardiac echocardiography (ICE) affords real time imaging of the cardiac structures during mapping and ablation. Methods: Between February and May 2007, 15 pa...

متن کامل

Fluoroless Catheter Ablation of Cardiac Arrhythmias: A 5-Year Experience.

BACKGROUND Catheter ablations have been traditionally performed with the use of fluoroscopic guidance, which exposes the patient and staff to the inherent risks of radiation. We have developed techniques to eliminate the use of fluoroscopy during cardiac ablations and have been performing completely fluoroless catheter ablations on our patients for over 5 years. METHODS We present a retrospec...

متن کامل

Nonfluoroscopic Ablation of Atrial Fibrillation Using Cryoballoon.

BACKGROUND The conventional method of cryoballoon ablation of atrial fibrillation involves the use of fluoroscopy for visual guidance. The use of fluoroscopy is accompanied by significant radiation risks to the patient and the medical staff. Herein, we report our experience in performing successful nonfluoroscopic pulmonary vein isolation using cryoballoon ablation in 5 consecutive patients wit...

متن کامل

Role of Intracardiac echocardiography in Atrial Fibrillation Ablation.

In the recent years, several new evidences support catheter-based ablation as a treatment modality of atrial fibrillation (AF). Based on a plenty of different applications, intracardiac echocardiography (ICE) is now a well-established technology in complex electrophysiological procedures, in particular in AF ablation. ICE contributes to improve the efficacy and safety of such procedures definin...

متن کامل

Recent Trends in Imaging for Atrial Fibrillation Ablation

Catheter ablation provides an important treatment option for patients with both paroxysmal and persistent atrial fibrillation. It mainly involves pulmonary vein isolation and additional ablations in the left atrium in persistent cases. There have been significant advancements in this procedure to enhance the safety and effectiveness. One of them is the evolution of various imaging modalities to...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Circulation. Arrhythmia and electrophysiology

دوره 2 6  شماره 

صفحات  -

تاریخ انتشار 2009