Confirmation of a false-positive result in CA 125 immunoradiometric assay caused by human anti-idiotypic immunoglobulin.
نویسندگان
چکیده
An immunoradiometric assay (IRMA) involving a monoclonal antibody (MAb OC125) to an ovarian carcinoma-associated antigenic determinant (CA 125) has been tested as one component in a strategy for early detection of epithelial ovarian cancer. We characterized one confirmed "false-positive" sample by murine antibody blocking studies, Western blotting, immunoaffinity, size-exclusion chromatography, and reactivity with polyclonal rabbit antisera to CA 125 antigen. The positive response of this serum in the CA 125 IRMA was due to a human IgM. The discrepant IgM was isolated from the serum by successive immunoaffinity steps with nonspecific murine MAb, MAb OC125, and goat antibodies to human IgM Fc. Purified IgM inhibited the binding of MAb OC125 to CA 125. Furthermore, rabbit antisera to CA 125 antigen competitively inhibited the binding of MAb OC125 to both CA 125 and the discrepant IgM. The discrepant activity thus appears to reflect binding of this human IgM to a idiotope of MAb OC125. Radioiodination of MAb OC125 by a different technique eliminated the discrepant activity and decreased the incidence of CA 125 positivity in an at-risk population of apparently healthy women, increasing the specificity of the IRMA to 99.8% in this group.
منابع مشابه
Evidence for interaction of human anti-idiotypic antibodies with CA 125 determination in a patient after radioimmunodetection.
Very high concentrations of CA 125 have been found in some ovarian cancer patients after repeated radioimmunodetection with anti-CA 125 antibodies [OC125-F(ab')2]. In one patient we measured a CA 125 concentration of 135,000 kilo-arb. units/L, using an enzyme immunoassay involving OC125 antibodies. With an immunoradiometric assay involving use of two new anti-CA 125 antibodies (B43.13 and B27.1...
متن کاملInterference by human anti-mouse antibodies in CA 125 assay after immunoscintigraphy: anti-idiotypic antibodies not neutralized by mouse IgG but removed by chromatography.
Falsely increased concentrations of the ovarian carcinoma-associated antigen, CA 125, were measured by a monoclonal antibody (MAb)-based double determinant immunoradiometric assay (IRMA) in patients who developed antibodies to mouse immunoglobulins (IgGs) after receiving injections of the same MAb as is used in the CA 125 IRMA. Addition of undiluted mouse serum or purified mouse IgG to the assa...
متن کاملHeteroantibody: Phantom of the Immunoassay
The heterophilic antibody, or heteroantibody, has been haunting immunoassays for many years. Interference by heteroantibodies in human sera can either increase or decrease the results of an immunoassay. For instance, by binding to a site other than the analyte-binding site, the heteroantibodies are capable of cross-linking the signal antibody with the capture antibody, thereby generating a fals...
متن کاملThe potency of immunoglobulin G fragments for inhibition of interference caused by anti-immunoglobulin antibodies in a monoclonal immunoradiometric assay for thyrotropin.
Heterophile antibodies in patients' serum may produce false increases in apparent analyte concentrations in "sandwich"-type immunoassays. Using three patients with endogenous anti-mouse IgG antibodies and a two-site mouse monoclonal assay for thyrotropin, we studied the ability of IgG fragments to block this positive interference. Mouse whole IgG and IgG Fc fragment blocked the interference vir...
متن کاملMonoclonal Antibody Immunoradiometric Assay for an Antigenic Determinant (CA 125) Associated with Human Epithelial Ovarian Carcinomas1
CA 125 is an antigenic determinant expressed by greater than 80% of nonmucinous epithelial ovarian carcinomas. An immunoradiometric assay has been developed using a murine mono clonal antibody (OC125) to quantitate CA 125 in human serum. This immunoradiometric assay was optimized for specificity, sensitivity, and performance characteristics. Using a simultane ous immunoradiometric assay, the me...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 34 6 شماره
صفحات -
تاریخ انتشار 1988