Effect of protein concentration on the determination of digoxin in serum by fluorescence polarization immunoassay.
نویسندگان
چکیده
Determination of digoxin by fluorescence polarization immunoassay (FPIA) with the Abbott "TDx" is significantly influenced by the concentration of total serum protein. Each 10 g/L increase in serum protein results in an 8% decrease in measured digoxin. Studies with [3H]digoxin confirmed that digoxin binds to the protein pellet during the trichloroacetic acid precipitation step before the immunoassay. Serum protein, or equal concentrations of albumin or gamma-globulin, exert an equivalent effect on the apparent digoxin value. Because the total protein concentration of the assay calibrators is low (50 g/L) compared with its reference interval in serum (60-80 g/L), results by FPIA may be expected to be low by an average of 16% (range, 8-24%). Digoxin results by FPIA will be most nearly accurate when the calibrators include a total protein concentration of about 70 g/L. Patients' specimens with abnormally high or low protein content will give falsely high or low results for digoxin.
منابع مشابه
Positive and negative interference of the Chinese medicine Chan Su in serum digoxin measurement. Elimination of interference by using a monoclonal chemiluminescent digoxin assay or monitoring free digoxin concentration.
An over-the-counter Chinese medicine, Chan Su, is used as a cardiotonic agent. We demonstrated significant digoxin-like immunoreactivity in various organic and aqueous extracts of Chan Su. For example, when a 20-microL aliquot of an aqueous extract of Chan Su powder (1 mg/mL) was added to a 2-mL aliquot of a drug-free serum, the observed digoxin-like immunoreactivity was 2.76 ng/mL (3.53 nmol/L...
متن کاملDigoxin-like immunoreactivity eliminated from serum by centrifugal ultrafiltration before fluorescence polarization immunoassay of digoxin.
Digoxin determined in the Abbott "TDx" by fluorescence polarization immunoassay by the manufacturer's recommended method involving precipitation of protein with 5-sulfosalicylic acid (SSA) is subject to interference from endogenous compounds having digoxin-like immunoreactivity. Guided by the work of Graves et al. (Clin Chem 1986;32:1506-9), we eliminated interference caused by digoxin-like imm...
متن کاملCross-reactivity of anti-digoxin antibodies with digitoxin depends on tracer structure.
The most common methods for measuring digoxin concentrations in serum are immunoassays. The prerequisite for exact determination of the digoxin value is an antibody that specifically binds digoxin. Because digitoxin differs from digoxin only in the C-12 hydroxy group, it is difficult to obtain anti-digoxin antibodies that do not cross-react with this compound. During the development of a fluore...
متن کاملEffect of Indian Ayurvedic medicine Ashwagandha on measurement of serum digoxin and 11 commonly monitored drugs using immunoassays: study of protein binding and interaction with Digibind.
CONTEXT Ashwagandha, a popular Ayurvedic medicine, is now available in the United States. Alkaloids found in this herb have structural similarity with digoxin. OBJECTIVE To study potential interference of Ashwagandha with serum digoxin measurement by immunoassays. Potential interference was also investigated with immunoassays for 11 other commonly monitored drugs. In addition, interaction of ...
متن کاملEffect of Asian and Siberian ginseng on serum digoxin measurement by five digoxin immunoassays. Significant variation in digoxin-like immunoreactivity among commercial ginsengs.
Asian and Siberian ginsengs contain glycosides with structural similarities to digoxin. We studied potential interference of ginseng in 5 digoxin immunoassays in 3 Asian (2 liquid extracts, 1 capsule) and 3 Siberian ginseng preparations (1 liquid extract, 2 capsules). With the fluorescence polarization immunoassay (FPIA), we observed apparent digoxin activity in 1 Asian liquid preparation and i...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 30 11 شماره
صفحات -
تاریخ انتشار 1984