Automated dibucaine number measurement with DuPont Dimension ES and AR analyzers.
نویسندگان
چکیده
We report a fully automated method for determining dibucaine number (DN) in a single-run procedure involving Dimension cholinesterase (CHE) Flex pseudo-(P)CHE reagents. The method was developed and optimized with the "open channels" and "kinetic" software facilities of the Dimension-ES instrument, where the DN is calculated automatically by an algorithm from the ratio of the uninhibited and inhibited rates, measured bichromatically, from a single analysis. The protocol was satisfactorily assessed for substrate depletion, linearity, reagent stability. and the effects of different dibucaine concentrations. Validation was performed across a range of CHE activities (1.5-22 kU/L) representing the three main genotypes, UU, UA, and AA. The respective DNs (mean +/- SD), determined on the Dimension-ES, were 82.0 +/- 1.6 (n = 32), 71.0 +/- 3.1 (n = 10), and 23.0 +/- 2.7 (n = 14), with corresponding imprecisions (CV) of 0.3%, 0.6%, and 5.2% (intraassay) and 0.7%, 0.7%, and 8.6% (interassay). Comparisons with reference (x) laboratory values and the DuPont aca (x') procedure (n = 53) gave regression equations of: y = 0.88x + 11.2, r = 0.99, and y = 0.85x' + 11.9, r = 0.99. A separate trial conducted with a Dimension-AR instrument gave similar performances. We conclude that the new DN method is fast, efficient, and appropriate for clinical use.
منابع مشابه
Modification of fully automated total iron-binding capacity (TIBC) assay in serum and comparison with dimension TIBC method.
BACKGROUND We previously reported the development of a fully automated assay for total iron-binding capacity (TIBC) in serum, using a multipurpose automated analyzer. However, this method requires four different reagents and is thus useful only with a limited number of available analyzers. We simplified our original assay and compared the analytical performance of the modified method with that ...
متن کاملRELATIONSHIP BETWEEN SUCCINYLCHOLINEINDUCED APNEA AND DIBUCAINE NUMBER IN PATIENTS WITH GENETIC DEFICIENCY OF PLASMA CHOLINESTERASE
In this study, a commercially available kit in which propionylthiocholine is used as a substrate and a dibucaine solution to determine the variants of the enzyme were utilized. Plasma cholinesterase activity of 151 patients who underwent different operations was measured during a period of six months. The mean±SD of the enzyme activity in 150 patients was 4.39±1.17 kU/L .. Seventeen patie...
متن کاملPeykan check, a simple continuous quality control method for hematology analyzers
A simple inexpensive con!nuous quality control method, by means of eight basic blood coun!ng parameters, obtained from automated hematology analyzers, using pa!ent samples, is described. A few samples with low, normal and high values were selected and introduced to the instrument early in the morning as the first count, the results of which are plo#ed on the appropriate chart as dots. The sa...
متن کاملQuality Control in Clinical Laboratories
1.1 The automated analyzers in clinical laboratories Nowadays, the overwhelming majority of laboratory results in clinical laboratories is being generated by automated analyzers. Modern automated analyzers are highly sophisticated instruments which can produce a tremendous number of laboratory results in a very short time. This is achieved thanks to the integration of technologies from three di...
متن کاملLocal Anesthetic Binding to Thylakoid Membranes. Relation to Inhibition of Light-Induced Membrane Energization and Photophosphorylation
The association o f the lipophilic tertiary amine and local anesthetic dibucaine with osmotically shocked chloroplasts o f Spinacia oleracea L. cv. Monatol was investigated. Dibucaine, known as an effective inhibitor o f thylakoid membrane energization and ATP synthesis, ex hibited three distinct binding classes with chloroplasts: partitioning in the lipid phase o f the membranes, electrostati...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Clinical chemistry
دوره 41 5 شماره
صفحات -
تاریخ انتشار 1995