Prediction of the Dynamic Response of the Potentiometric Carbon Dioxide Electrode
نویسندگان
چکیده
The dynamic behaviour of the potentiometric carbon dioxide electrode is predicted by means of a digital simulation that considers both diffusion and reaction kinetics. The method allows for complete description of the electrode response without requiring the assumption that one of these processes can be neglected. The validity of the simulation was tested experimentally by using a commercial pC0, electrode provided with various silicone rubber and teflon membranes. The measured values are in good agreement with those of the simulation.
منابع مشابه
A Novel Carbon Paste Electrode for Potentiometric Determination of Vanadyl Ion
Carbon paste electrode was modified by disodium 2,4-diamino-7-methylbenzo[g] pteridine-5,10-dioxide to fabricate a sensor for potentiometric determination of vanadyl ion (VO2+) in aqueous medium. The new ligand was synthesized in three steps with high yield. The optimal composition of the carbon paste was made of graphite powder (64.5%), ionophore (16.6%), paraffin oil (17.8%), a...
متن کاملIon-Selective Carbon Paste Electrode Based on 2-Amino-N-benzthioazolyl Benzamide (ABTB) for Determination of Copper (II) by Potentiometric Method
2-Amino-N-benzthioazolyl benzamide (ABTB) was synthesized, characterized and used for thefabrication of a potentiometric sensor for Cu2+ metal ions. The electrode exhibits linear response to Cu(II) over a wide concentration range (4.79×10 -8 – 1.85×10 -1 M) with Nernstian slope of 30 ± 1.5 mVper decade. The electrode can be used in the pH range from 2 to 9. It has a fast response time of about1...
متن کاملSelective Potentiometric Sensor for Isoniazid Ultra-Trace Determination Based on Fe3O4 Nanoparticles Modified Carbon Paste Electrode (Fe3O4/CPE)
Significant effects of magnetic iron oxide (Fe3O4) nanoparticles on the potentiometric behavior of carbon paste electrode for bio determination of Isoniazid aromatic drug (INH) were studied. The electrochemical performance of modified Fe3O4/CPE was investigated. The constructed Fe3O4/CPE showed high precision, sensitivity ...
متن کاملDevelopment of novel potentiometric sensors for determination of lidocaine hydrochloride in pharmaceutical preparations, serum and urine samples
This article is focused on the determination of lidocaine hydrochloride as a local anaesthetic drug. A potentiometric method based on modified screen-printed and modified carbon paste ion-selective electrodes was described for the determination of lidocaine hydrochloride in different pharmaceutical preparations. It was based on potentiometric titration of lidocaine hydrochloride using modified ...
متن کاملDevelopment of novel potentiometric sensors for determination of lidocaine hydrochloride in pharmaceutical preparations, serum and urine samples
This article is focused on the determination of lidocaine hydrochloride as a local anaesthetic drug. A potentiometric method based on modified screen-printed and modified carbon paste ion-selective electrodes was described for the determination of lidocaine hydrochloride in different pharmaceutical preparations. It was based on potentiometric titration of lidocaine hydrochloride using modified ...
متن کامل