Effective dose estimation for oncological and neurological PET/CT procedures
نویسندگان
چکیده
BACKGROUND The aim of this study was to retrospectively evaluate the patient effective dose (ED) for different PET/CT procedures performed with a variety of PET radiopharmaceutical compounds. PET/CT studies of 210 patients were reviewed including Torso (n = 123), Whole body (WB) (n = 36), Head and Neck Tumor (HNT) (n = 10), and Brain (n = 41) protocols with 18FDG (n = 170), 11C-CHOL (n = 10), 18FDOPA (n = 10), 11C-MET (n = 10), and 18F-florbetapir (n = 10). ED was calculated using conversion factors applied to the radiotracer activity and to the CT dose-length product. RESULTS Total ED (mean ± SD) for Torso-11C-CHOL, Torso-18FDG, WB-18FDG, and HNT-18FDG protocols were 13.5 ± 2.2, 16.5 ± 4.5, 20.0 ± 5.6, and 15.4 ± 2.8 mSv, respectively, where CT represented 77, 62, 69, and 63% of the protocol ED, respectively. For 18FDG, 18FDOPA, 11C-MET, and 18F-florbetapir brain PET/CT studies, ED values (mean ± SD) were 6.4 ± 0.6, 4.6 ± 0.4, 5.2 ± 0.5, and 9.1 ± 0.4 mSv, respectively, and the corresponding CT contributions were 11, 14, 23, and 26%, respectively. In 18FDG PET/CT, variations in scan length and arm position produced significant differences in CT ED (p < 0.01). For dual-time-point imaging, the CT ED (mean ± SD) for the delayed scan was 3.8 ± 1.5 mSv. CONCLUSIONS The mean ED for body and brain PET/CT protocols with different radiopharmaceuticals ranged between 4.6 and 20.0 mSv. The major contributor to total ED for body protocols is CT, whereas for brain studies, it is the PET radiopharmaceutical.
منابع مشابه
Uncertainties in effective dose estimation for CT transmission scan in total body PET-CT imaging with Auto mA3D tube current modulation
Background: Positron Emission Tomography-Computed Tomography (PET-CT) is a useful hybrid imaging modality in the diagnosis of various malignancies. This modality imposes almost 20 mSv radiation dose to the patient. The purpose of the present study was to evaluate the uncertainties in calculated CT effective dose in TUBE CURRENT MODULATION-activated scans by Impact-Dose software. Materials and ...
متن کاملOptimization of oncological ¹⁸F-FDG PET/CT imaging based on a multiparameter analysis.
PURPOSE This paper describes a method to achieve consistent clinical image quality in (18)F-FDG scans accounting for patient habitus, dose regimen, image acquisition, and processing techniques. METHODS Oncological PET/CT scan data for 58 subjects were evaluated retrospectively to derive analytical curves that predict image quality. Patient noise equivalent count rate and coefficient of variat...
متن کاملEstimation of fetal absorbed dose from low-dose attenuation-correction CT in PET / CT imaging by using the Body Builder Phantom
One of the methods of studying the physiology and metabolism of important tissues such as the heart, brain and cancer tumors is the use of PET/CT System. A small number of patients are pregnant women who undergo a PET/CT scan due to lack of knowledge about pregnancy or due to dire clinical need. The final dose received by the fetus is based on three factors: the absorbed dose of the fetal tissu...
متن کاملEvaluation of Radiation Exposure to Staff and Environment Dose from [18F]-FDG in PET/CT and Cyclotron Center using Thermoluminescent Dosimetry
Background: PET/CT imaging using [18F]-FDG is utilized in clinical oncology for tumor detecting, staging and responding to therapy procedures. Essential consideration must be taken for radiation staff due to high gamma radiation in PET/CT and cyclotron center. The aim of this study was to assess the staff exposure regarding whole body and organ dose and to evaluate environment dose in PET/CT an...
متن کاملEstablishment of Diagnostic Reference Levels and Estimation of Effective Dose from Computed Tomography Head Scans at a Tertiary Hospital in South Africa
Introduction: Head scans are the most frequently performed computed tomography (CT) examinations worldwide. However, there is growing concern over the probability of increased cancer risks among the exposed populations. Diagnostic reference levels (DRLs) identify radiation dose that is not commensurate with clinical objectives. The aim of this study was to establish DR...