The influence of stem length and fixation on initial femoral component stability in revision total knee replacement

نویسندگان

  • N. Conlisk
  • H. Gray
  • P. Pankaj
  • C. R. Howie
چکیده

OBJECTIVES Orthopaedic surgeons use stems in revision knee surgery to obtain stability when metaphyseal bone is missing. No consensus exists regarding stem size or method of fixation. This in vitro study investigated the influence of stem length and method of fixation on the pattern and level of relative motion at the bone-implant interface at a range of functional flexion angles. METHODS A custom test rig using differential variable reluctance transducers (DVRTs) was developed to record all translational and rotational motions at the bone-implant interface. Composite femurs were used. These were secured to permit variation in flexion angle from 0° to 90°. Cyclic loads were applied through a tibial component based on three peaks corresponding to 0°, 10° and 20° flexion from a normal walking cycle. Three different femoral components were investigated in this study for cementless and cemented interface conditions. RESULTS Relative motions were found to increase with flexion angle. Stemmed implants reduced relative motions in comparison to stemless implants for uncemented constructs. Relative motions for cemented implants were reduced to one-third of their equivalent uncemented constructs. CONCLUSIONS Stems are not necessary for cemented implants when the metaphyseal bone is intact. Short cemented femoral stems confer as much stability as long uncemented stems.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Coupling failure between stem and femoral component in a constrained revision total knee arthroplasty.

Knee revision using constrained implants is associated with greater stresses on the implant and interface surfaces. The present report describes a case of failure of the screw coupling between the stem and the femoral component. The cause of the failure is surmised with outline of the treatment in this case with extensive femoral bone loss. Revision implant stability was augmented with the use ...

متن کامل

Tapered modular fluted titanium stems for femoral fixation in revision total knee arthroplasty

Consensus regarding femoral stem fixation options in revision total knee arthroplasty remains controversial. Tapered, modular, fluted titanium (TMFT) stems have an excellent track record in total hip arthroplasty for their ability to provide axial and rotational stability in situations of compromised host bone. We present 3 successfully treated cases in which the Food & Drug Administration gran...

متن کامل

Interprosthetic femoral fractures: analysis of 14 cases. Proposal for an additional grade in the Vancouver and SoFCOT classifications.

INTRODUCTION Interprosthetic fracture is a rare but serious entity, impairing consolidation and stability due to adverse mechanical conditions related to bone fragility and implant volume. OBJECTIVE The present study highlights the difficulties involved in managing such fractures, details treatment options and reports findings leading to a proposed additional grade in the comparable Vancouver...

متن کامل

The long-term results of press-fit cemented stems in total knee prostheses.

We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in migration and rotation, was obtained by pressurised antibioti...

متن کامل

Hip Arthroplasty and its Revision in a Child: Case Report and Literature Review

Juvenile idiopathic arthritis is the leading cause of hip replacement in young children. However, arthroplasty in this population is challenging with several concerns about quality of the growing bone, young age for revision surgery, and difficulties in potential several revisions. In this study we introduce a case of a 12-year old who is one of the youngest patients to undergo revision hip art...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 1  شماره 

صفحات  -

تاریخ انتشار 2012