A methodology for on‐board CBCT imaging dose using optically stimulated luminescence detectors

نویسندگان

  • Noor Mail
  • Muhammad Yusuf
  • Nazeeh Alothmany
  • A. Abdulrahman Kinsara
  • Fahad Abdulkhaliq
  • Suliman M. Ghamdi
  • Abdelhamid Saoudi
چکیده

Cone-beam computed tomography CBCT systems are used in radiation therapy for patient alignment and positioning. The CBCT imaging procedure for patient setup adds substantial radiation dose to patient's normal tissue. This study pre-sents a complete procedure for the CBCT dosimetry using the InLight optically-stimulated-luminescence (OSL) nanoDots. We report five dose parameters: the mean slice dose (DMSD); the cone beam dose index (CBDIW); the mean volume dose (DMVD); point-dose profile, D(FOV); and the off-field Dose. In addition, CBCT skin doses for seven pelvic tumor patients are reported. CBCT-dose mea-surement was performed on a custom-made cylindrical acrylic body phantom (50cm length, 32cm diameter). We machined 25 circular disks (2 cm thick) with grooves and holes to hold OSL-nanoDots. OSLs that showed similar sensitivities were selected and calibrated against a Farmer-type ionization-chamber (0.6 CT) before being inserted into the grooves and holes. For the phantom scan, a standard CBCT-imaging protocol (pelvic sites: 125 kVp, 80 mA and 25 ms) was used. Five dose parameters were quantified: DMSD, CBDIW, DMVD, D(FOV), and the off-field dose. The DMSD for the central slice was 31.1 ± 0.85 mGy, and CBDIW was 34.5± 0.6 mGy at 16cm FOV. The DMVD was 25.6 ± 1.1 mGy. The off-field dose was 10.5 mGy. For patients, the anterior and lateral skin doses attributable to CBCT imaging were 39.04 ± 4.4 and 27.1 ± 1.3 mGy, respectively.OSL nanoDots were convenient to use in measuring CBCT dose. The method of selecting the nanoDots greatly reduced uncertainty in the OSL measurements. Our detailed calibration procedure and CBCT dose measurements and calculations could prove useful in developing OSL routines for CBCT quality assessment, which in turn gives them the property of high spatial resolution, meaning that they have the potential for measurement of dose in regions of severe dose-gradients.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Measurement of eye lens dose for Varian On-Board Imaging with different cone-beam computed tomography acquisition techniques

The objective of this work was to measure patient eye lens dose for different cone-beam computed tomography (CBCT) acquisition protocols of Varian's On-Board Imaging (OBI) system using optically stimulated luminescence dosimeter (OSLD) and to study the variation in eye lens dose with patient geometry and distance of isocenter to the eye lens. During the experimental measurements, OSLD was place...

متن کامل

Measurement of skin dose from cone-beam computed tomography imaging

OBJECTIVE To measure surface skin dose from various cone-beam computed tomography (CBCT) scanners using point-dosimeters. MATERIALS & METHODS A head anthropomorphic phantom was used with nanoDOT optically stimulated luminescence (OSL) dosimeters (Landauer Corp., Glenwood, IL) attached to various anatomic landmarks. The phantom was scanned using multiple exposure protocols for craniofacial eva...

متن کامل

An optically stimulated luminescence dosimeter for measuring patient exposure from imaging guidance procedures.

There is a growing interest in patient exposure resulting from an x-ray imaging procedure used in image-guided radiation therapy. This study explores a feasibility to use a commercially available optically stimulated luminescence (OSL) dosimeter, nanoDot, for estimating imaging radiation exposure to patients. The kilovoltage x-ray sources used for kV-cone-beam CT (CBCT) imaging acquisition proc...

متن کامل

The effective dose of different scanning protocols using the Sirona GALILEOS(®) comfort CBCT scanner.

OBJECTIVES To determine the effective dose and CT dose index (CTDI) for a range of imaging protocols using the Sirona GALILEOS(®) Comfort CBCT scanner (Sirona Dental Systems GmbH, Bensheim, Germany). METHODS Calibrated optically stimulated luminescence dosemeters were placed at 26 sites in the head and neck of a modified RANDO(®) phantom (The Phantom Laboratory, Greenwich, NY). Effective dose...

متن کامل

Dosimetric Analysis of Microscopic Disease in SBRT for Lung Cancers

OBJECTIVE The objective of this study is to theoretically and experimentally evaluate the dosimetry in the microscopic disease regions surrounding the tumor under stereotactic body radiation therapy of lung cancer. METHODS For simplicity, the tumor was considered moving along 1 dimension with a periodic function. The probability distribution function of the tumor position was generated accord...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2016