Bench-test comparison of 26 emergency and transport ventilators
نویسندگان
چکیده
INTRODUCTION Numerous emergency and transport ventilators are commercialized and new generations arise constantly. The aim of this study was to evaluate a large panel of ventilators to allow clinicians to choose a device, taking into account their specificities of use. METHODS This experimental bench-test took into account general characteristics and technical performances. Performances were assessed under different levels of FIO2 (100%, 50% or Air-Mix), respiratory mechanics (compliance 30,70,120 mL/cmH2O; resistance 5,10,20 cmH2O/mL/s), and levels of leaks (3.5 to 12.5 L/min), using a test lung. RESULTS In total 26 emergency and transport ventilators were analyzed and classified into four categories (ICU-like, n = 5; Sophisticated, n = 10; Simple, n = 9; Mass-casualty and military, n = 2). Oxygen consumption (7.1 to 15.8 L/min at FIO2 100%) and the Air-Mix mode (FIO2 45 to 86%) differed from one device to the other. Triggering performance was heterogeneous, but several sophisticated ventilators depicted triggering capabilities as efficient as ICU-like ventilators. Pressurization was not adequate for all devices. At baseline, all the ventilators were able to synchronize, but with variations among respiratory conditions. Leak compensation in most ICU-like and 4/10 sophisticated devices was able to correct at least partially for system leaks, but with variations among ventilators. CONCLUSION Major differences were observed between devices and categories, either in terms of general characteristics or technical reliability, across the spectrum of operation. Huge variability of tidal volume delivery with some devices in response to modifications in respiratory mechanics and FIO2 should make clinicians question their use in the clinical setting.
منابع مشابه
Performance comparison of 15 transport ventilators.
BACKGROUND Numerous mechanical ventilators are designed and marketed for use in patient transport. The complexity of these ventilators differs considerably, but very few data exist to compare their operational capabilities. METHODS Using bench and animal models, we studied 15 currently available transport ventilators with regard to their physical characteristics, gas consumption (duration of ...
متن کاملEvaluation of transport ventilators at mild simulated altitude: a bench study in a hypobaric chamber.
BACKGROUND Previous studies on ventilators used for air transport showed significant effects of altitude, in particular with regard to accuracy of the tidal volume (VT) and breathing frequency. The aim of the study was to evaluate transport ventilators under hypobaric conditions. METHODS We conducted a bench study of 6 transport ventilators in a Comex hypobaric chamber to simulate mild altitu...
متن کاملNear-infrared spectroscopy with vascular occlusion test may not be the adequate tool to explore microcirculation in pulmonary arterial hypertension.
essentials of respiratory care, 4th edition. St. Louis:Mosby Elsevier; 2005:103. 13. L’Her E, Roy A. Bench tests of simple, handy ventilators for pandemics: performance, autonomy, and ergonomy. Respir Care 2011;56(6):751-760. 14. Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC. Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventila...
متن کاملEvaluation of ventilators used during transport of critically ill patients: a bench study.
OBJECTIVE To evaluate the most recent transport ventilators' operational performance regarding volume delivery in controlled mode, trigger function, and the quality of pressurization in pressure support mode. METHODS Eight recent transport ventilators were included in a bench study in order to evaluate their accuracy to deliver a set tidal volume under normal resistance and compliance conditi...
متن کاملAn overnight comparison of two ventilators used in the treatment of chronic respiratory failure.
Differences between bilevel ventilators used for noninvasive intermittent positive pressure ventilation (NIPPV) have been demonstrated during bench testing. However, there are no clinical studies comparing these machines. The authors have previously shown that the Quantum pressure support ventilator and Sullivan variable positive airway pressure II ST differ in performance during bench testing....