Comparison of crossed screw versus plate fixation for radial neck fractures.
نویسندگان
چکیده
BACKGROUND Fixation of radial neck fractures can be achieved with a plate and screw construct or, in absence of comminution, with two obliquely-oriented screws. This study investigated the mechanical properties, specifically the stiffness and load to failure, of these two fixation strategies in a cadaver model. METHODS Ten matched-pair radii were removed from fresh cadaver arms. A transverse osteotomy was created at the neck of each radius. Right-sided radii were fixed with two oblique headless compression screws; left-sided radii were fixed with a radial neck plate. The distal aspect of each radius was potted in urethane casting resin. The radial head was loaded in shear in 4 different planes (medial to lateral, lateral to medial, posterior to anterior, and anterior to posterior) using an Instron machine. Stiffness and load to failure were recorded. FINDINGS The stiffness of both constructs was similar in all planes except for loading from medial to lateral where the screw construct was 1.8 times stiffer. Average ultimate failure occurred at 229N for the screws and 206N for the plate. Failure strength was not statistically different. However, mode of failure differed for both fixation constructs, the plate failed in bending while the screws failed by pullout and fracture. INTERPRETATION The two strategies provide similar strength and stiffness for the fixation of transverse, non-comminuted radial neck fractures. While plate and screw constructs are more appropriate for axially unstable or comminuted fractures, two oblique screws might be preferred for simple transverse neck fractures since this strategy requires less exposure and the implant is buried.
منابع مشابه
Comparison of three different fixation constructs for radial neck fractures: a biomechanical study
BACKGROUND Fixation of radial neck fractures can be achieved with a plate and screw construct or with two screws. This study evaluated the biomechanical properties of three different fixation methods following radial neck fractures. METHODS Twenty-four fourth-generation composite radii were sawed to simulate an unstable radial neck fracture. They were then instrumented with a plate and screw ...
متن کاملScrew fixation versus arthroplasty versus plate fixation for 3-part radial head fractures.
PURPOSE To compare the outcome following headless compression screw fixation versus radial head arthroplasty versus plate fixation for 3-part Mason types III or IV radial head fracture. METHODS Records of 25 men and 16 women aged 21 to 80 (mean, 43.3) years who underwent fixation using 2 to 3 2-mm cannulated headless compression screws (n=16), radial head arthroplasty (n=13), or fixation with...
متن کاملFirst Metatarsophalangeal Joint Arthrodesis: A Retrospective Comparison of Crossed-screws, Locking and Non-Locking Plate Fixation with Lag Screw
Background:Locking plate fixation is increasingly used for first metatarsophalangeal joint (MTP-I) arthrodesis. Still there is few comparable clinical data regarding this procedure. Methods:We retrospectively evaluated 60 patients who received an arthrodesis of the MTP-I between January 2008 and June 2010. With 20 patients each we performed a locking plate fixation with lag screw, arthrodesis w...
متن کاملFirst Metatarsophalangeal Joint Arthrodesis: A Retrospective Comparison of Crossed-screws, Locking and Non-Locking Plate Fixation with Lag Screw
Background:Locking plate fixation is increasingly used for first metatarsophalangeal joint (MTP-I) arthrodesis. Still there is few comparable clinical data regarding this procedure. Methods:We retrospectively evaluated 60 patients who received an arthrodesis of the MTP-I between January 2008 and June 2010. With 20 patients each we performed a locking plate fixation with lag screw, arthrodesis w...
متن کاملA Pitfall in Fixation of Distal Humeral Fractures with Pre-Contoured Locking Compression Plate
Anatomically precontoured locking plates are intended to facilitate the fixation of articular fractures and particularly those associated with osteoporosis. Fractures of the distal humerus are relatively uncommon injuries where operative intervention can be exceptionally challenging. The distal humeral trochlea provides a very narrow anatomical window through which to pass a fixed-angle locking...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical biomechanics
دوره 30 9 شماره
صفحات -
تاریخ انتشار 2015