Reel syndrome: a rare cause of pacemaker malfunction.

نویسندگان

  • Luis Alvarez-Acosta
  • Rafael Romero Garrido
  • Marcos Farrais-Villalba
  • Julio Hernández Afonso
چکیده

To cite: Alvarez-Acosta L, Romero Garrido R, FarraisVillalba M, et al. BMJ Case Rep Published online: [please include Day Month Year] doi:10.1136/bcr-2014204545 DESCRIPTION An 82-year-old woman was admitted to our hospital because of syncope. A 12-lead ECG demonstrated atrial fibrillation with a ventricular response of 35 bpm, and a VVIR (ventricular pacing, ventricular sensing, inhibiting mode, rate response function) pacemaker was implanted using a ventricular active fixation lead via the left subclavian artery. Prior to discharge a chest X-ray was taken and showed normal ventricular lead placement but with a minor lead retraction (figure 1A). One month after implantation she came again to the emergency room with dizziness with a ventricular response of 42 bpm. A chest X-ray was urgently performed and showed ventricular lead retraction (figure 1B, black arrow) secondary to rotation of the pulse generator on its transverse axis with subsequent coiling of the lead (figure 1B, white arrow). This situation has been named ‘Reel Syndrome’ and it happens when the generator rotates around its sagittal axis, causing the electrode to roll up like a spool above or below the generator. The patient underwent emergent ventricular lead repositioning and the pulse generator was fixed to the pectoral muscle without any further complications. We postulate that the minor lead retraction that was noticed with the first chest X-ray should have warned us about the incomplete fixation of the lead to the pectoral muscle. Reel syndrome is a rare cause of an implantable device malfunctioning. It is commonly included in the macrodislocation lead-dysfunctioning syndromes along with Twiddler’s and Ratchet syndromes. They differ from each other in the causing mechanism. Twiddler’s syndrome is caused by retraction and dislocation of the electrodes due to rotation of the generator around the axis defined by the electrode. Although an external manipulation by the patient can facilitate it, this would not be a necessary condition. Owing to this rotational movement, the electrode winds as a braid, which defines the characteristic appearance of this finding. Ratchet syndrome is caused by retraction and electrode dislocation with ratcheting but without coiling of the generator due to progressive displacement of the electrodes from their fixing protections. Twiddler’s and Reel syndromes have similar aetiologies; female gender, large pocket, obesity, children, older people and dementia can be listed as contributing factors and their prevalence is unknown. Reel syndrome commonly occurs within a month of implantation and normally there is no damage of the leads. This is the reason why normally there is no need of lead change, unlike Twiddler’s syndrome where the leads are normally damaged and their replacement is usually mandatory. Chest X-ray is a simpler and better method for diagnosis and therefore it is always requested when such a complication is suspected because it can easily differentiate between these three syndromes (figure 2).

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Reel Syndrome--A Variety of Twiddler's Syndrome.

Reel syndrome is an uncommon condition in which a patient's subconscious, inadvertent, or deliberate external manipulation of their cardiac pacemaker results in the movement of the transducer wires along its transverse axis. We report a clinical case of a 67 year old lady with permanent pacemaker presenting with pacemaker malfunction and was detected with reel syndrome.

متن کامل

Subclavian crush syndrome: a cause of pacemaker lead fracture.

A 48-year-old male with a symptomatic 2:1 atrio-ventricular block and a dual-chamber pacemaker, implanted one year previously, was admitted due to a syncopal episode. Pacemaker malfunction was identified as the cause of syncope. Subclavian crush syndrome was the cause of the pacemaker malfunction. Its incidence, consequences and management are discussed in this report.

متن کامل

Pectoral muscular twitching: a rare manifestation of spontaneous twiddler syndrome

The twiddler syndrome is an uncommon cause of pacemaker malfunction, in which twisting or rotating of the device in its pocket results in lead retraction or coiling and subsequent malfunction of an implanted devices such as a pacemaker, an implantable cardioverter-defibrillator (ICD), cardiac resynchronization therapy (CRT).  This report describes a rare case of twiddler syndrome after implant...

متن کامل

Unusual lead helix damage due to ‘Selective’ Reel syndrome

Oblivious manipulation of the device by the patient is referred to as "Twiddler's" syndrome. It is characterized by the coiling of the pacemaker lead due to the rotation of the pacemaker generator on its long axis. However, the rotation of the pacemaker generator on its transverse axis with subsequent coiling of the pacemaker leads around the pulse generator is called as ''Reel syndrome''. In t...

متن کامل