Effects of Titanium Mesh Cage End Structures on the Compressive Load at the Endplate Interface: A Cadaveric Biomechanical Study
نویسندگان
چکیده
BACKGROUND This study aimed to evaluate whether obliquely angled and ring-shaped titanium mesh cage (TMC) end structures can improve the compressive load on the endplate interface in anterior cervical corpectomy and fusion (ACCF). MATERIAL AND METHODS A total of 23 volunteers underwent cervical lateral x-ray. The oblique angle of the superior endplate was measured, which was used to construct the gradient of the TMC end. Forty-two fresh cadaveric vertebral bodies were harvested and randomly distributed among four TMC groups with different ends. The baseline indicators of bone mineral density and anteroposterior and transverse dimensions were recorded. The superior endplate was placed at an angle of 12° when performing uniaxial compression testing. The maximum loads of the four TMCs were assessed. RESULTS There were no significant differences among the groups regarding the baseline indicators. The conventional TMC had the lowest maximum load (1362.3±221.78 N, p<0.05), whereas the TMC with an obliquely end ring had the highest maximum load (2095.82±285.64 N, p<0.05). The maximum loads of the TMCs with oblique footprints and flat end ring were much higher than that of the conventional TMC (p<0.05) but significantly lower than that of the TMC with the obliquely end ring (p<0.05), with average values of 1806.91±246.98 N and 1725.3±213.33 N, respectively. CONCLUSIONS Both the ring shape and oblique angle of the TMC end contributed to an increase in compressive force and are advocated for use in TMC structure optimization to decrease the incidence of TMC subsidence in ACCF.
منابع مشابه
Effect of follower load on motion and stiffness of the human thoracic spine with intact rib cage.
Researchers have reported on the importance of the rib cage in maintaining mechanical stability in the thoracic spine and on the validity of a compressive follower preload. However, dynamic mechanical testing using both the rib cage and follower load has never been studied. An in vitro biomechanical study of human cadaveric thoracic specimens with rib cage intact in lateral bending, flexion/ext...
متن کاملEffects on inadvertent endplate fracture following lateral cage placement on range of motion and indirect spine decompression in lumbar spine fusion constructs: A cadaveric study
BACKGROUND The lateral transpsoas approach to interbody fusion is gaining popularity. Existing literature suggests that perioperative vertebra-related complications include endplate breach owing to aggressive enedplate preparation and poor bone quality. The acute effects of cage subsidence on stabilization and indirect decompression at the affected level are unknown. The purpose of this study w...
متن کاملCorpectomy cage subsidence with rectangular versus round endcaps.
Corpectomy cages with rectangular endcaps utilize the stronger peripheral part of the endplate, potentially decreasing subsidence risk. The authors evaluated cage subsidence during cyclic biomechanical testing, comparing rectangular versus round endcaps. Fourteen cadaveric spinal segments (T12-L2) were dissected and potted at T12 and L2, then assigned to a rectangular (n=7) or round (n=7) endca...
متن کاملThe Stabilising Effect of the Anterior Oblique Ligament to Prevent Directional Subluxation at the Trapeziometacarpal Joint of the Thumb: A Biomechanical Cadaveric Study
Background: The trapeziometacarpal joint (TMCJ) is inherently unstable, relying on ligament restraint to preventsubluxation. Subluxation of the thumb in a dorsoradial direction is often observed in clinical practice, either after acuteligament injury or more commonly with osteoarthritis (OA). This subluxation follows loss of function of trapeziometacarpalligaments that stabilise this joint, res...
متن کاملAn FE investigation simulating intra-operative corrective forces applied to correct scoliosis deformity
BACKGROUND Adolescent idiopathic scoliosis (AIS) is a deformity of the spine, which may require surgical correction by attaching a rod to the patient's spine using screws implanted in the vertebral bodies. Surgeons achieve an intra-operative reduction in the deformity by applying compressive forces across the intervertebral disc spaces while they secure the rod to the vertebra. We were interest...
متن کامل