Clinical evaluation of the accuracy and precision of the CDI 500 in-line blood gas monitor with and without gas calibration.

نویسندگان

  • Anne Louise Bellaiche
  • Peter F Nielsen
  • Steven Brantlov
  • Marianne B Møller
  • Michael Winterdahl
چکیده

During cardiopulmonary bypass blood gases can be analyzed with laboratory equipment or with an in-line monitor giving instant results. The manufacturer of the CDI 500 in-line blood gas monitor recommends gas calibration before use. In acute cases there may not be time to perform a gas calibration. We hypothesized that after calibration against laboratory results, the CDI values of pH, pO2, and pCO2 will keep the same level of accuracy, whether the CDI has been gas calibrated or not. We performed a prospective randomized observational study using a study group without gas calibration (29 patients) and a control group with gas calibration (29 patients). Blood sampling was done at the beginning of bypass, and 30 minutes later. After each blood sample the CDI was in-vivo calibrated to the values simultaneously obtained from the ABL. Before in-vivo calibration values from the CDI without gas calibration were significantly different from the ABL-values in accuracy as well as precision, whereas the results from the gas calibrated CDI were largely consistent with the ABL. Before in-vivo calibration, the CDI without gas calibration was completely unreliable. After in-vivo calibration there was no statistical difference between the values of the CDI with and without calibration. We recommend gas calibration of the CDI before use in the period before in-vivo calibration.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Clinical evaluation of a new in-line continuous blood gas monitor.

Two methodologies for obtaining accurate blood gas and electrolyte values during cardiopulmonary bypass (CPB) are traditional laboratory analyzers, which use an electrochemical technology, and continuous in-line monitoring systems, which use a fluorometric and/or spectrophotometric technology. The purpose of the present study was to evaluate the accuracy of a new continuous in-line monitor, the...

متن کامل

Assessment of a continuous blood gas monitoring system in animals during circulatory stress

BACKGROUND The study was aimed to determine the measurement accuracy of The CDI™ blood parameter monitoring system 500 (Terumo Cardiovascular Systems Corporation, Ann Arbor MI) in the real-time continuous measurement of arterial blood gases under different cardiocirculatory stress conditions METHODS Inotropic stimulation (Dobutamine 2.5 and 5 μg/kg/min), vasoconstriction (Arginine-vasopressin...

متن کامل

Use of the CDI blood parameter monitoring system 500 for continuous blood gas measurement during extracorporeal membrane oxygenation simulation.

UNLABELLED The Oximetrix III Opticath (Abbott Critical Care Systems) is used for continuous measurement of venous saturation in a variety of applications, including extracorporeal membrane oxygenation (ECMO), despite clinical reports that have presented data showing poor accuracy of these devices. The CDI Blood Parameter Monitoring System 500 (Terumo) is an inline blood gas monitoring tool comm...

متن کامل

Gelsomino 1 Assessment of a Continuous Blood Gas Monitoring System in Animals during Circulatory stress. Short Title: Continuous Blood Gas Monitoring

Introduction: The study was aimed to determine the measurement accuracy of The CDI TM blood parameter monitoring system 500 (Terumo Cardiovascular Systems Corporation, Ann Arbor MI) in the real-time continuous measurement of arterial blood gases under different cardiocirculatory stress conditions Methods: Inotropic stimulation (Dobutamine 2.5 and 5 μg/kg/min), vasoconstriction (Argininevasopres...

متن کامل

An Improved GC Method for Rapid Analysis of Valproic Acid in Human Plasma Without Derivatization

        A simple, reproducible and rapid gas chromatographic method for precise determination of valproic acid (VPA) in human plasma has been developed. Total time for sample preparation and GC analysis is less than 45 min. After plasma protein precipitation, VPA was extracted into chloroform with suitable recovery. By using Stabilwax®<span styl...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The journal of extra-corporeal technology

دوره 43 2  شماره 

صفحات  -

تاریخ انتشار 2011