Evaluation of a tapered polymer-free sirolimus-eluting stent in porcine coronary arteries

نویسندگان

  • Qiu-Ping Shi
  • Bo Zheng
  • Ming Chen
  • Xin-Gang Wang
  • Bin Zhang
چکیده

Background: No stent is yet available for tapering coronary arteries. The purpose of the present study is to evaluate the applicability, efficacy and safety of a novel tapered, polymer-free sirolimus eluting stent (t-PFSES) specifically designed for tapering coronary arteries. Methods: 28 pigs underwent the placement of 56 oversized stents (control, n = 28, t-PFSES, n = 28), and quantitative coronary angiography (QCA) and histopathologic analysis were performed at 1, 3 and 6 months follow-up. Results: In proximal segments, both the t-PFSES and control stents lead to similar results, while, in distal segments, the t-PFSES stent was associated with significantly improved angiographic outcomes. Compared with the control stents, t-PFSES exhibited a greater inhibition of neointimal hyperplasia in all stented segments, but the magnitude of the neointimal area was lower in the proximal segments than in the distal segments, (1.15 vs. 1.48 mm2, 0.33 vs. 0.8 mm2, 0.3 vs. 1.14 mm2 at 1, 3 and 6 months, respectively, P > 0.05). Complete re-endothelialization was observed with both stents at 1 month post-procedure, but the t-PFSES stented distal segments had numerically lower inflammation scores (P = NS). Conclusion: The t-PFSES stent is applicable to coronary artery segments with marked tapering (10% tapering or 0.45 mm). Compared with the conventional bare metal stent, t-PFSES appeared to be safer and lead to superior angiographic outcomes, especially in the distal segments of tapering coronary arteries.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Polymer-free dual drug-eluting stents evaluated in a porcine model

BACKGROUND Although drug-eluting stents have dramatically reduced the rates of restenosis and target lesion revascularization, they are associated with stent thrombosis (ST), a catastrophic event likely due to delayed endothelialization and chronic inflammation caused by the polymer and the metal scaffolds. To increase the safety and efficacy of stents, polymer-free dual drug-eluting stents (DD...

متن کامل

Histopathological Analysis of Drug Eluting Stents in Porcine Coronary Restenosis Model

The aim of this study was to compare the two types of drug-eluting stents with histological analysis in porcine coronary restenosis model. Pigs were randomized into two groups in which the coronary arteries (12 pigs, 12 coronaries in each group) had either biolimus-eluting stent with biodegradable polymer (BPS) or sirolimus-eluting stent with durable polymer (DPS). Radiological analysis was per...

متن کامل

Inhibition of neointima formation by a novel drug-eluting stent system that allows for dose-adjustable, multiple, and on-site stent coating.

OBJECTIVE The risk of in-stent restenosis can be considerably reduced by stents eluting cytostatic compounds. We created a novel drug-eluting stent system that includes several new features in the rapidly evolving field of stent-based drug delivery. METHODS AND RESULTS The aim of the present study was the preclinical evaluation of a stent-coating system permitting individual, on-site coating ...

متن کامل

Nanoporous CREG-Eluting Stent Attenuates In-Stent Neointimal Formation in Porcine Coronary Arteries

BACKGROUND The goal of this study was to evaluate the efficacy of a nanoporous CREG-eluting stent (CREGES) in inhibiting neointimal formation in a porcine coronary model. METHODS In vitro proliferation assays were performed using isolated human endothelial and smooth muscle cells to investigate the cell-specific pharmacokinetic effects of CREG and sirolimus. We implanted CREGES, control sirol...

متن کامل

Vascular response profiles following a nano polymer-free sirolimus-eluting stent implantation assessed by optical coherence tomography in a porcine model

Optical coherence tomography (OCT) is a high resolution imaging modality and has been considered as the ideal tool for the evaluation of neointimal tissue and vascular responses following stent implantation. However, vascular response profiles following the implantation of a novel nano polymer-free sirolimus-eluting stent (SES) assessed by OCT has not been fully investigated. Therefore, the aim...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2016