Stability of serum ferritin measured by immunoturbidimetric assay after storage at -80°C for several years

نویسندگان

  • Anne-Sylvia Sacri
  • Daniela Ferreira
  • Babak Khoshnood
  • Laurent Gouya
  • Henrique Barros
  • Martin Chalumeau
چکیده

BACKGROUND Iron deficiency (ID) may impair long-term neurological development when it occurs in young infants. In cohort studies, it is sometimes necessary to evaluate ID with sera kept frozen for several years. To assess ID, learned societies recommend measuring serum ferritin (SF) level combined with C-reactive protein level. The long-term stability of C-reactive protein in frozen samples is well established but not ferritin. METHODS We measured SF level (immunoturbidimetric assay; in micrograms per liter) immediately after collection from 53 young adults recruited and followed-up in Porto, Portugal, from 2011 to 2013 (SF1), and then, in 2016 in two aliquots kept frozen at- 80°C for 3 to 5 years: one without (SF2A) and one with (SF2B) intermediate thawing in 2014. We compared SF1 to SF2A then SF2B; statistical agreement was evaluated by the Bland and Altman method and the effect of intermediate thawing by regression modelling. RESULTS Mean SF2A-SF1 and SF2B-SF1 differences were -2.1 (SD 7.0) and 48.9 (SD 66.9). Values for Bland and Altman 95% limits of agreement were higher for the comparison of SF2B and SF1 than SF2A and SF1: -82.2 to 179.9 and -15.8 to 11.8, respectively; the effect of thawing was highly significant (p <0.001). CONCLUSIONS Agreement between SF values before and after 3 to 5 years of constant freezing at -80°C was in a generally accepted range, which supports the hypothesis of ferritin's stability at this temperature for a long period. In long-term storage by freezing, intermediate thawing induced a major increase in values.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Validation of a commercially available automated canine-specific immunoturbidimetric method for measuring canine C-reactive protein

BACKGROUND Measurement of C-reactive protein (CRP) is used for diagnosing and monitoring systemic inflammatory disease in canine patients. An automated human immunoturbidimetric assay has been validated for measuring canine CRP, but cross-reactivity with canine CRP is unpredictable. OBJECTIVE The purpose of the study was to validate a new automated canine-specific immunoturbidimetric CRP meth...

متن کامل

The Relation between Left Ventricular Function and Serum Ferritin in Major Β-Thalassemia

Abstract Objective Cardiac dysfunction is a major cause of death in patients with beta thalassemia. In these patients, repeated blood transfusion, ineffective erythropoiesis and increased gastrointestinal iron absorption lead to iron overload in the body and this induced heart failure. Left ventricular ejection fraction was measured in major beta thalassemia (β-Th) patients to detect the rela...

متن کامل

Flow Cytometric Measurement of CD41/CD61 and CD42b Platelet Receptors and Clotting Assay of Platelet Factor 3 During Long Term-Storage of Platelet Concentrates

Background: The purpose of the present in vitro study was to evaluate the effect of long term storage of conventional platelet concentrates (PCs) on major platelet receptors CD42b and CD41/CD61 by flow cytometry method and also measuring the overall platelet procoagulant activity status using platelet factor 3 (PF3) assay. Materials and Methods: Six random units of conventional platelet conce...

متن کامل

A new High Sensitive Functional Nephelometrical Assay for Assaying C- reactive protein in Serum Based on Phosphocholine Interaction

C-reactive protein (CRP) is able to bind phosphocholine in the presence of calcium ions. According to a previous functional property of CRP, we tried to develop an affordable and cheap high sensitive nephelometric CRP assay using soy oil. Serum samples were measured by Nephelometer BNII (Siemens), by mixing the serum with diluted soy oil emulsion (Intralipid ® 20%) and Tris-calcium buffer (PH 7...

متن کامل

Association between serum ferritin level and diastolic cardiac function in patients with major β-thalassemia

Abstract Background Prevention of myocardial siderosis is a key step to reduce rate of mortality in thalassemic patients. Our objective was to study association between echocardiography parameters and serum ferritin level in patients with major thalassemia. Materials and Methods Sixty-six patients with major thalassemia were studied in Amir Kabir hospital, Arak, Iran. Serum ferritin levels ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2017