Long‐term evaluation and cross‐checking of two geometric calibrations of kV and MV imaging systems for Linacs

نویسندگان

  • Tsuicheng D. Chiu
  • Yulong Yan
  • Ryan Foster
  • Weihua Mao
چکیده

Geometric or mechanical accuracy of kV and MV imaging systems of two Varian TrueBeam linacs have been monitored by two geomertirc calibration systems, Varian IsoCal geometric calibration system and home-developed gQA system. Results of both systems are cross-checked and the long-term geometric stabilities of linacs are evaluated. Two geometric calibration methodologies have been used to assess kV and MV imaging systems and their coincidence periodically on two TrueBeam linacs for about one year. Both systems analyze kV or MV projection images of special designed phantoms to retrieve geometric parameters of the imaging systems. The isocenters — laser isocenter and centers of rotations of kV imager and EPID — are then calculated, based on results of multiple projections from different angles. Long-term calibration results from both systems are compared for cross-checking. There are 24 sessions of side-by-side calibrations performed by both systems on two TrueBeam linacs. All the disagreements of isocenters between two calibrations systems are less than 1 mm with ± 0.1 mm SD. Most of the large disagreements occurred in vertical direction (AP direction), with an averaged disagreement of 0.45 mm. The average disagreements of isocenters are 0.09 mm in other directions. Additional to long-term calibration monitoring, for the accuracy test, special tests were performed by misaligning QA phantoms on purpose (5 mm away from setup isocenter in AP, SI, and lateral directions) to test the liability performance of both systems with the known deviations. The errors are within 0.5 mm. Both geometric calibration systems, IsoCal and gQA, are capable of detecting geometric deviations of kV and MV imaging systems of linacs. The long-term evaluation also shows that the deviations of geometric parameters and the geometric accuracies of both linacs are small and very consistent during the one-year study period.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of IsoCal geometric calibration system for Varian linacs equipped with on‐board imager and electronic portal imaging device imaging systems

The purpose of this study is to evaluate the accuracy and reproducibility of the IsoCal geometric calibration system for kilovoltage (kV) and megavoltage (MV) imagers on Varian C-series linear accelerators (linacs). IsoCal calibration starts by imaging a phantom and collimator plate using MV images with different collimator angles, as well as MV and kV images at different gantry angles. The sof...

متن کامل

Evaluation of the Machine Performance Check application for TrueBeam Linac

BACKGROUND Machine Performance Check (MPC) is an application to verify geometry and beam performances of TrueBeam Linacs, through automated checks based on their kV-MV imaging systems. In this study, preliminary tests with MPC were analyzed using all photon beam energies of our TrueBeam, comparing whenever possible with external independent checks. METHODS Data acquisition comprises a series ...

متن کامل

Comparative Study between Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) for Radiation Treatment Verifications

Introduction: Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) are the preferred tools of Image Guided Radiotherapy (IGRT) and Dose Guided Radiotherapy (DGRT) which have been used for Radiotherapy treatment verifications. As a result, the number of publications dealing with these two tools for radiation treatment verification has increased consi...

متن کامل

Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments

A clinical workflow was developed for urgent palliative radiotherapy treatments that integrates patient simulation, planning, quality assurance, and treatment in one 30-minute session. This has been successfully tested and implemented clinically on a linac with MV CBCT capabilities. To make this approach available to all clin-ics equipped with common imaging systems, dose calculation accuracy b...

متن کامل

A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for Varian's 2D/2D (kV/MV) and 3D/3D (planCT/CBCT) matching was developed. The procedure was to check: (1) the coincidence of on-board imager (OBI), portal imager (PI), and cone beam CT (CBCT)'s isocenters (digital graticules) to a linac's isocenter (to a pre-specified accuracy); (2) that the positioning difference detected by 2D/2D (kV/MV) and 3D/3D(planCT/CBCT) matc...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2015