Dose modification factor analysis of multilumen balloon brachytherapy applicator with Monte Carlo simulation
نویسندگان
چکیده
The Contura brachytherapy applicator is a silicone balloon with five lumens in which a high-dose-rate brachytherapy source can traverse. Multilumen applicators, like the Contura, are used in accelerated partial breast irradiation (APBI) brachytherapy in instances where asymmetric dose distributions are desired; for example, when the applicator surface-to-skin thickness is small (< 7 mm). In these instances, the air outside the patient and the lung act as a poor scattering medium, scattering less dose back into the breast and affecting the dose distribution. The recent report by Task Group 186 of the American Association of Physicists in Medicine (AAPM) has outlined the importance of moving towards brachytherapy dose calculations using heterogeneity corrections. However, at this time, many commercial treatment planning systems do not correct for tissue heterogeneity, which can result in inaccuracies in the planned dose distribution. To quantify the deviation in the skin dose we utilize the dose modification factor (DMF), defined as the ratio of the dose rate at 1 cm beyond the applicator surface with homogenous medium, to the dose rate at 1 cm with heterogeneous medium. This investigation models the Contura applicator with the Monte Carlo N-Particle code version 5, and determines a DMF through simulation. Taking all geometrical considerations into account, an accurate model of the Contura balloon applicator was created in MCNP and used to run simulations. The dose modification factor was found to be only slightly dependent on whether the dose distribution was symmetric or asymmetric. These results indicate that the dose delivered to part of the PTV may be lower than the planned dose by up to 12%, and that these brachytherapy plans should be viewed with caution. In addition to studying the effects of backscatter, an evaluation was made regarding the capabilities of the Contura device to shape an asymmetric dose distribution. We compared these results to a previous study of a MammoSite ML and a SAVI device and found that the dose shaping capabilities of the Contura were quite similar to that of the MammoSite ML, but markedly inferior to the SAVI.
منابع مشابه
Proof of Symmetric Dose Distribution in Mammosite Applicator in Breast Brachytherapy with MCNP Simulation
Introduction: In brachytherapy treatments, high-dose sources are used in interstitial placement. In this paper, the modeling of a common breast cancer treatment applicator called MammoSite is done by the Monte Carlo simulation code. Then, to study and calculate the amount of doses in the breast and the dose rate of the organs at risk of radiation, including: lungs, ribs and ski...
متن کاملCalculation of absorbed dose rate in the lungs, ribs and skin by the mammosite applicator in breast cancer brachytherapy with MCNP
Introduction: Breast cancer is the most common cancer among women. Brachytherapy is one of the ways to treat breast cancer. In brachytherapy treatments, high-dose sources are used in interstitial placement. Here, Iridium-192 is used. This source emits gamma rays with energies ranging from 136 kV to 1060 kV electron volts. The activity of the used spring is 10 Ci. The simulated ...
متن کاملDosimetry of 188Re and 186Re sources based on Monte Carlo calculations for endovascular brachytherapy after balloon angioplasty
Background: Recent pre-clinical and clinical studies indicate that irradiation in the dose range of 15 to 30 Gy can reduce rate of restenosis in patients who have undergone an angioplasty. The use of filled balloon with radioactive solution was proposed as one of the possible intravascular irradiation techniques. Materials and Methods: The Monte Carlo N-particle Transport Code (MCNP4b) was used...
متن کاملDosimetric characteristics of 137Cs sources used in after loading Selectron system by Monte Carlo method
Background: For an effective treatment planning in brachytherapy, it is necessary to know the accurate source dosimetric information such as air kerma strength, exposure rate constant, dose rate constant and redial dose distribution. The usual method to determine these factors is thermo luminescent dosimeter (TLD) dosimetry. Nowadays, another more accurate method is known to be the Monte Carlo ...
متن کاملDose modification factors for 192Ir high‐dose‐rate irradiation using Monte Carlo simulations
A recently introduced brachytherapy system for partial breast irradiation, MammoSite, consists of a balloon applicator filled with contrast solution and a catheter for insertion of an 192Ir high-dose-rate (HDR) source. In using this system, the treatment dose is typically prescribed to be delivered 1 cm from the balloon's surface. Most treatment-planning systems currently in use for brachythera...