AbstractID: 8312 Title: The use of dual-energy CT images for Monte Carlo treatment planning: material extraction and metal artifact reduction
ثبت نشده
چکیده
ID: 8312 Title: The use of dual-energy CT images for Monte Carlo treatment planning: material extraction and metal artifact reduction Purpose: To investigate the feasibility of dual-energy CT-based material extraction (DECT) for Monte Carlo (MC) dose calculations, to evaluate metal streaking artifact reduction using DECT and to propose a novel scanning technique for motion artifacts reduction. Method and Materials: A phantom with 17 tissue equivalent materials was scanned at two tube voltages. The accuracy of mass density (ρ) extraction using DECT (ρdual) compared to the standard single-energy CT technique (ρsingle) and the accuracy of the effective atomic number (Z) extraction was evaluated. A comparison of ρsingle and ρdual for a canine subject was made. The DECT extraction was also used on a prostate phantom and a patient with brachytherapy seeds in order to evaluate metal streaking artifact reduction. Moreover, a novel scanning technique with a moving copper filter is proposed. An optimal filter thickness is found and a MC simulation of DECT is performed. Results: The mean error on the relative electron density (ρe) and Z extraction is 1.8% and 3.2%. The improvement in ρdual compared to the standard ρsingle is significant, especially in soft bone materials where the ρ assignment error decreases from 8-10% to less than 2% when the DECT technique is used. Similar conclusions are drawn from the canine study where the comparison showed large differences (up to 12%) in bone marrow. The metal streaking artifacts caused by the brachytherapy seeds in the prostate phantom and the patient were significantly reduced. A 2 mm copper filter thickness was found to be optimal for the novel scanning technique. MC simulations of a CT scanner confirmed that with this filtration materials can be extracted with an accuracy of 1.8% in ρe and 3.6% in Z. Conclusion: The dual-energy CT technique results in improved tissue extraction and reduced metal artifacts, and is therefore a useful tool for MC dose calculations.
منابع مشابه
Metal Artifact Reduction of Dental Fillings in Head and Neck CT Images
Introduction: The issue of metal artifact and its reduction is as old as the clinical use of computed tomography itself. When metal objects such as dental fillings, hip prostheses or surgical clips are present in the computed tomography (CT) field of view (FOV), make severe artifacts that reduce the image quality and accuracy of CT numbers. They can lead to unreliable ...
متن کاملA New Method for Metal Artifact Reduction in CT Scan Images
Introduction In CT imaging, metallic implants inside the tissues cause metal artifact that reduce the quality of image for diagnosis. In order to reduce the effect of this artifact, a new method with more appropriate results has been presented in this research work. Materials and Methods The presented method comprised of following steps: a) image enhancement and metal areas extraction, b) sinog...
متن کاملExtracting Material Information from the CT Numbers by Artificial Neural Networks for Use in the Monte Carlo Simulations of Different Tissue Types in Brachytherapy
Background: The artificial neural networks (ANNs) are useful in solving nonlinear processes, without the need for mathematical models of the parameters. Since the relationship between the CT numbers and material compositions is not linear, ANN can be used for obtaining tissue density and composition.Objective: The aim of this study is to utilize ANN for determination of the composition and mass...
متن کاملCT-Based Brachytherapy Treatment Planning using Monte Carlo Simulation Aided by an Interface Software
Introduction: In brachytherapy, radioactive sources are placed close to the tumor, therefore, small changes in their positions can cause large changes in the dose distribution. This emphasizes the need for computerized treatment planning. The usual method for treatment planning of cervix brachytherapy uses conventional radiographs in the Manchester system. Nowadays, because of their advantages ...
متن کاملEvaluation of Metal Artifact Reduction software in Computed Tomography
Introduction: The image quality of computed tomography (CT) can be seriously lowered by metal implants of patients. These implants are known to exert a significant impact on diagnostic accuracy due to artifacts. The current study aimed to assess the usefulness of Metal Artifact Reduction (MAR) software in the reduction of metal artifacts, in comparison to iterative rec...
متن کامل