Comparison of 45-degree Kelman and 45-degree balanced phaco tip designs in torsional microcoaxial phacoemulsification.

نویسندگان

  • Süleyman Demircan
  • Mustafa Ataş
  • Emre Göktaş
  • Burhan Başkan
چکیده

AIM To compare the intraoperative performance and postoperative outcome after microcoaxial torsional phacoemulsification using either a Kelman or balanced phaco tip. METHODS Cataracts were treated using 2.2 mm microcoaxial torsional phacoemulsification using either a 45-degree mini-flared Kelman(®) or a 45-degree Intrepid(®) Balanced phaco tip. Intraoperative measurements included total ultrasound (US) time, cumulative dissipated energy (CDE), torsional US time, and balanced salt solution (BSS) use. The central endothelial cell density (ECD) and central corneal thickness (CCT) were evaluated preoperatively and postoperatively 1, 7, and 30d after surgery using noncontact specular microscopy. RESULTS The 116 enrolled eyes (116 patients) were divided equally between the Kelman and balanced tip groups. Intraoperative measurements showed significantly less total US time, torsional US time, CDE, and BSS use in the balanced group than in Kelman group (P<0.05). The total US time, torsional US time, CDE, and BSS use were 17.45±14.53s, 16.63±13.97s, 6.38±5.26, and 48.21±17.21 mL in the Kelman group and 11.39 ± 9.60s, 10.90 ± 9.25s, 4.04 ± 3.42, and 41.36 ± 12.70 mL in the balanced group, respectively. CONCLUSION Torsional phacoemulsification performed with a balanced tip provided more effective lens removal with less total US time, torsional time, CDE, and BSS use, as well as similar changes in ECD with a Kelman tip in all cataract grades. This special designed phaco tip for torsional phacoemulsification provides an alternative phaco tip for many surgeons' preference with straight phaco tip.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Relationship between endothelial cell loss and microcoaxial phacoemulsification parameters in noncomplicated cataract surgery

PURPOSE To assess the relationship between postoperative endothelial cell loss and microcoaxial phaco parameters using Ozil IP (Alcon Laboratories, Inc, Fort Worth, TX) in noncomplicated cataract surgery. METHODS In this prospective observational study, 120 consecutive cases of cataract patients with different grades of nuclear hardness underwent microcoaxial phacoemulsification through a 2.2...

متن کامل

Effect of IOP based infusion system with and without balanced phaco tip on cumulative dissipated energy and estimated fluid usage in comparison to gravity fed infusion in torsional phacoemulsification

BACKGROUND To evaluate the effect of three different combinations of tip designs and infusion systems in torsional phacoemulsification (INFINITI and CENTURION) in patients with cataract. According to the manufacturer, two unique improvements in the Centurion are: active fluid dynamic management system and use of an intrepid balanced tip. The study specifically aimed to evaluate the beneficial e...

متن کامل

Laboratory assessment of thermal characteristics of three phacoemulsification tip designs operated using torsional ultrasound

PURPOSE Ultrasound activation of phacoemulsification (phaco) tips can create considerable thermal energy that may increase the risk of tissue damage during cataract surgery. The purpose of this study was to define the thermal profiles of three phaco tip designs in simulated surgical conditions. METHODS In this laboratory investigation, sleeved phaco tips (mini-flared Kelman(®) tip with aspira...

متن کامل

Heat Generation and Efficiency of a New Modified Phaco Tip and Sleeve

PURPOSE To compare a modified phacoemulsification tip with the established micro tip, in terms of temperature at the corneal wound, efficiency, and chatter events, using the Centurion® Vision system. METHODS Eighty porcine eyes were randomized into 4 groups: 1)sleeveless conventional 45D MiniFlared ABS® Kelman tip (1.1-mm incision); 2)sleeveless new modified 45D ABS® INTREPID® balanced tip(1....

متن کامل

Combined 30-degree bevel up and down technique against 0-degree phaco tip for phacoemulsification surgery of hard cataracts

PURPOSE To evaluate the effective phaco time (EPT), cumulative dissipated energy (CDE) and nucleus emulsification time (NET) as phaco parameters with 0- and 30-degree phaco tip. PATIENTS AND METHODS This prospective, interventional and observational case series included 294 patients scheduled for cataract removal by phacoemulsification technique. Patients with nuclear cataracts of grade 4 and...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • International journal of ophthalmology

دوره 8 6  شماره 

صفحات  -

تاریخ انتشار 2015