The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants

نویسندگان

  • Ali Fayaz
  • Minoo Mahshid
  • Aboulfazl Saboury
  • Seyed Jalil Sadr
  • Ghassem Ansari
چکیده

BACKGROUND Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices (F-S MTLDs) in achieving their target torque values. MATERIALS AND METHODS Peak torque measurements of 15 new F-S MTLDs from three different manufacturers (Astra Tech, BioHorizons, Dr. Idhe) were measured ten times before and after 100 steam sterilization using a digital torque gauge. To simulate the clinical situation of aging (number of use) target torque application process was repeated 10 times after each sterilization cycle and the peak torque values were registered. Comparison of the mean differences with target torque in each cycle was performed using one sample t test. Considering the type of MTLDs as inter subject comparison, One-way repeated measure ANOVA was used to evaluate the absolute values of differences between devices of each manufacturer in each group (α = 0.05). RESULTS The results of this study in Dr. Idhe group showed that, mean of difference values significantly differed from the target torque (P = 0.002) until 75 cycles. In Astra Tech group, also mean of difference values with under estimation trend, showed a significant difference with the target torque (P < 0.001). Mean of difference values significantly differed from the target torque with under estimation trend during all the 100 cycles in BioHorizons group (P < 0.05). CONCLUSION The torque output of each individual device stayed in 10% difference from target torque values before 100 sterilization cycles, but more than 10% difference from the target torque was seen in varying degrees during these consequent cycles.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of Component Dismantling on Quality of Sterilization of Mechanical Torque Limiting Devices(Pilot Study)

Objectives Dental instruments may serve as a route of infection transmission. The necessity of dismantling of Friction-Style (FS) and SpringStyle (SS) mechanical torque limiting devices (MTLDs) prior to disinfection and sterilization is a matter of debate, aside fromthe fact that it is difcult and time-consuming. This study sought to assess the effect of dismantling of components on quality of ...

متن کامل

Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants

OBJECTIVE Accurate delivery of torque to implant screws is critical to generate ideal preload in the screw joint and to offer protection against screw loosening. Mechanical torque-limiting devices (MTLDs) are available for this reason. In this study, the accuracy of one type of friction-style and two types of spring-style MTLDs at baseline, following fatigue conditions and sterilization process...

متن کامل

The effect of steam sterilization on the accuracy of spring-style mechanical torque devices for dental implants

BACKGROUND Mechanical torque devices (MTDs) are one of the most commonly recommended devices used to deliver optimal torque to the screw of dental implants. Recently, high variability has been reported about the accuracy of spring-style mechanical torque devices (S-S MTDs). Joint stability and survival rate of fixed implant supported prosthesis depends on the accuracy of these devices. Currentl...

متن کامل

Accuracy of mechanical torque-limiting devices for dental implants after clinical service.

STATEMENT OF PROBLEM Limited information is available regarding the accuracy of mechanical torque-limiting devices (MTLDs) after their clinical use. PURPOSE The purpose of this study was to determine the accuracy of 2 types of MTLDs (friction-style and spring-style) after clinical use. MATERIAL AND METHODS Twenty-seven MTLDs in clinical service at The Ohio State University College of Dentis...

متن کامل

The Effect of Aging on the Accuracy of New Friction-Style Mechanical Torque Limiting Devices for Dental Implants

OBJECTIVE High variability in delivering the target torque is reported for friction-style mechanical torque limiting devices (F-S MTLDs). The effect of aging (number of use) on the accuracy of these devices is not clear. The purpose of this study was to assess the effect of aging on the accuracy (±10% of the target torque) of F-S MTLDs. MATERIALS AND METHODS Fifteen new F-S MTLDs and their ap...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2014