Which Phantom Is Better for Assessing the Image Quality in Full-Field Digital Mammography?: American College of Radiology Accreditation Phantom versus Digital Mammography Accreditation Phantom
نویسندگان
چکیده
OBJECTIVE To compare between the American College of Radiology (ACR) accreditation phantom and digital mammography accreditation phantom in assessing the image quality in full-field digital mammography (FFDM). MATERIALS AND METHODS In each week throughout the 42-week study, we obtained phantom images using both the ACR accreditation phantom and the digital mammography accreditation phantom, and a total of 42 pairs of images were included in this study. We assessed the signal-to-noise ratio (SNR) in each phantom image. A radiologist drew a square-shaped region of interest on the phantom and then the mean value of the SNR and the standard deviation were automatically provided on a monitor. SNR was calculated by an equation, measured mean value of SNR-constant coefficient of FFDM/standard deviation. Two breast radiologists scored visible objects (fibers, specks, and masses) with soft-copy images and calculated the visible rate (number of visible objects/total number of objects). We compared SNR and the visible rate of objects between the two phantoms and calculated the k-coefficient for interobserver agreement. RESULTS The SNR of the ACR accreditation phantom ranged from 42.0 to 52.9 (Mean, 47.3 ± 2.79) and that of Digital Phantom ranged from 24.8 to 54.0 (Mean, 44.1 ± 9.93) (p = 0.028). The visible rates of all three types of objects were much higher in the ACR accreditation phantom than those in the digital mammography accreditation phantom (p < 0.05). Interobserver agreement for visible rates of objects on phantom images was fair to moderate agreement (k-coefficients: 0.34-0.57). CONCLUSION The ACR accreditation phantom is superior to the digital mammography accreditation phantom in terms of SNR and visibility of phantom objects. Thus, ACR accreditation phantom appears to be satisfactory for assessing the image quality in FFDM.
منابع مشابه
Evaluation of Doses and Image Quality in Mammography with Screen-Film, CR, and DR Detectors – Application of the ACR Phantom
BACKGROUND Different methods of image quality evaluation are routinely used for analogue and digital mammography systems in Poland. In the present study, image quality for several screen-film (SF), computed radiography (CR), and fully digital (DR) mammography systems was compared directly with the use of the ACR mammography accreditation phantom. MATERIAL/METHODS Image quality and mean glandu...
متن کاملAssessment of mammography quality in Istanbul.
PURPOSE We aimed to evaluate the mammography image quality in İstanbul and to survey the awareness of mammography quality. MATERIALS AND METHODS Fifty-five mammography units in İstanbul were visited. The mammography image quality was evaluated using a standard American College of Radiology (ACR) accreditation phantom. The phantom contained 16 objects, and a score of at least 10 of 16 was acce...
متن کاملEvaluation of Various Mammography Phantoms for Image Quality Assessment in Digital Breast Tomosynthesis
We investigated the appropriateness of four different mammography phantoms for image quality evaluation in Digital Breast Tomosynthesis (DBT). We tested the CIRS BR3D phantom, the ACR Prototype FFDM Accreditation Phantom, the Penn anthropomorphic breast phantom and the Quart mam/digi EPQC phantom. This work discusses the advantages and shortcomings of each phantom and concludes that none of the...
متن کاملAutomatic Scoring of American College of Radiology (acr) Phantom Image Using Image Processing Technique By
A phantom image is used to perform quality control in mammographic facilities. The interpretation of the ACR phantom images is depend upon the experience, visual threshold, scoring criteria and state of the observer. This paper summarizes the first step of automating this score using digital image processing, namely template matching via cross correlation, and suggests a metric for the visibili...
متن کاملA phantom using titanium and Landolt rings for image quality evaluation in mammography.
A phantom for image quality evaluation of digital mammography systems is presented and compared to the most widely used phantoms in Europe and the US. The phantom contains objects for subjective detection of Landolt rings (four-alternative, forced-choice task) and for objective calculation of signal-difference-to-noise ratios (SDNR), both in a titanium background within a 12-step wedge. Evaluat...