Peak scatter factors for high energy photon beams.

نویسنده

  • X A Li
چکیده

The peak scatter factor (PSF) for a photon beam is defined as the ratio of the total dose and the primary dose at the depth of dose maximum. The values of the PSF for photon beams ranging from 60Co to 24 MV are calculated using the EGS4 Monte Carlo technique, to avoid measurement difficulties. The calculation shows that the effect of SSD on PSF for high energy photon beams is not significant for small fields, but can be as high as 1% for large fields. For the 60Co beam, the calculation agrees with the data tabulated on BJR Supplement 25 to within 0.8%. The BJR value (1.054) of 10 x 10 cm2 for 60Co is 0.6% lower than the present value due to the underestimation of scatters from the source capsule and collimators. For a given field size, PSF is varied by up to 2% when beam quality changes from 60Co to 24 MV. For normalized PSF, the values of BJR Supplement 25 (which are assumed to be the same for beams ranging from 60Co to 50 MV) agree well with the present calculation for small field sizes, but are higher than our data by up to 2% for large field sizes. The presently calculated PSFs are related to field size(s) by an empirical expression, PSF = 1 + ms/(s + n), where m and n are the fitting parameters. This equation describes the PSFs within 0.4% (0.15% on average).

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Calculating Weighting Factors for Mixing Megavoltage Photon Beams to Achieve Desirable Dose Distribution in Radiotherapy

Background: In radiotherapy, low-energy photon beams are better adapted to the treated volume, and the use of high-energy beams can reduce hot spots in the radiation therapy. Therefore, mixing low and high energies with different ratios can control the rate of hotspots, as well as the dose distribution of the target volume.Material and Methods: The percentage depth doses (PDDs) were calculated ...

متن کامل

Determination of the Energy Windows for the Triple Energy Window Scatter Correction Method in Gadolinium-159 Single Photon Emission Computed Tomography Using Monte Carlo Simulation

Introduction: In radionuclide imaging, object scatter is one of the major factors leading to image quality degradation. Therefore, the correction of scattered photons might have a great impact on improving the image quality. Regarding this, the present study aimed to determine the main and sub-energy windows for triple energy window (TEW) scatter correction method usin...

متن کامل

Dose Calculations for Lung Inhomogeneity in High-Energy Photon Beams and Small Beamlets: A Comparison between XiO and TiGRT Treatment Planning Systems and MCNPX Monte Carlo Code

Introduction Radiotherapy with small fields is used widely in newly developed techniques. Additionally, dose calculation accuracy of treatment planning systems in small fields plays a crucial role in treatment outcome. In the present study, dose calculation accuracy of two commercial treatment planning systems was evaluated against Monte Carlo method. Materials and Methods Siemens Once or linea...

متن کامل

Calculation the phantom scatter factor for the linear accelerators device

Introduction: In the last few decades, a lot of monte carlo codes have been introduced for medical applications. this is important because through simulation, we are fully independent of the spectral scattering factor that we are not able measure in real terms. The effect of the radiation energy, the dimension of the radiation field, the sensitive volume of the ion chamber on t...

متن کامل

Calculating weighting factors for mixing megavoltage photon beams to achieve desirable dose distribution in Radiotherapy

Introduction: Many studies have shown the effects of delivered dose distribution due to the incident photon energy on the tumor and healthy tissues. The ability to access the most appropriate radiation energy is essential to achieve the optimal treatment planning but there is a serious limitation in number of energies available on radiation therapy machines can restrict it. <s...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 26 6  شماره 

صفحات  -

تاریخ انتشار 1999