Dual‐axis rotational coronary angiography can reduce peak skin dose and scattered dose: a phantom study
نویسندگان
چکیده
The purpose of this study was to evaluate peak skin dose received by the patient and scattered dose to the operator during dual-axis rotational coronary angiography (DARCA), and to compare with those of standard coronary angiography (SA). An anthropomorphic phantom was used to simulate a patient undergoing diagnostic coronary angiography. Cine imaging was applied on the phantom for 2 s, 3 s, and 5 s in SA projections to mimic clinical situations with normal vessels, and uncomplicated and complicated coronary lesions. DARCA was performed in two curved trajectories around the phantom. During both SA and DARCA, peak skin dose was measured with thermoluminescent dosimeter arrays and scattered dose with a dosimeter at predefined height (approximately at the level of left eye) at the operator's location. Compared to SA, DARCA was found lower in both peak skin dose (range: 44%-82%, p < 0.001) and scattered dose (range: 40%-70%, p < 0.001). The maximal reductions were observed in the set mimicking complicated lesion examinations (82% reduction for peak skin dose, p < 0.001; 70% reduction for scattered dose, p < 0.001). DARCA reduces both peak skin dose and scattered dose in comparison to SA. The benefi t of radiation dose reduction could be especially signifi cant in complicated lesion examinations due to large reduction in X-ray exposure time. The use of DARCA could, therefore, be recommended in clinical practice to minimize radiation dose.
منابع مشابه
Breast Radiation Protection in Coronary Angiography
Introduction: Coronary angiography (CA) is an interventional diagnostic procedure to assess cardiovascular disease. CA is performed using dedicated fluoroscopic systems equipped with either flat detector or image intensifier. There is an increase in the number of CA annually. Reducing the radiation exposure in the CA is essential, especially due to high fluoroscopy time. In ord...
متن کاملAssessment of Maximum Entrance Skin Dose of Patients Undergoing Cardiac Interventional Procedures and Its Correlation with Other Dosimetric Parameters
Introduction: In recent years, the number of complex coronary angiography (CA) is increasing rapidly. These procedures have a significant contribution to medical exposure to the general population. Exposure of patients to high doses of x-rays could cause deterministic effects on the skin. Therefore, the assessment of radiation doses of patients is of great importance. ...
متن کامل3D cerebral angiography: radiation dose comparison with digital subtraction angiography.
BACKGROUND AND PURPOSE As the use of 3D rotational angiography (3D RA) for the evaluation of cerebral vasculature becomes more widespread, it is important to evaluate this imaging method's effect on patient radiation dose. The purpose of the study is to measure 3D RA radiation dose as compared with biplanar digital subtraction angiography (DSA). METHODS The distribution and peak skin dose wer...
متن کاملAssessment of patients’ entrance skin and effective dose in a mathematical human phantom for the most common interventional radiological examinations
Introduction: Nowadays, we are witnessing an exponential use of interventional radiology techniques in different communities. After CT, interventional techniques are the second factor of increasing patients’ doses in different societies. Measuring patient doses from aforementioned methods has been recommended by many radiation protection professional organizations such as ICRP...
متن کاملRadiation dose and physical image quality in 128‐section dual‐source computed tomographic coronary angiography: a phantom study
One-hundred-and-twenty-eight-section dual X-ray source computed tomography (CT) systems have been introduced into clinical practice and have been shown to increase temporal resolution. Higher temporal resolution allows low-dose spiral mode at a high pitch factor during CT coronary angiography. We evaluated radiation dose and physical image qualities in CT coronary angiography by applying high-p...
متن کامل