Closure of atrial septal defects with the Amplatzer occlusion device: preliminary results.
نویسندگان
چکیده
OBJECTIVES This study reports our clinical experience with transcatheter closure of secundum atrial septal defects (ASDs) in children, using the Amplatzer, a new occlusion device. BACKGROUND None of the devices previously used for transcatheter closure of interatrial communications has gained wide acceptance. METHODS We examined the efficacy and safety of the Amplatzer, a new self-centering septal occluder that consists of two round disks made of Nitinol wire mesh and linked together by a short connecting waist. Sixteen patients with secundum ASD met established two- and three-dimensional echocardiographic and cardiac catheterization criteria for transcatheter closure. The Amplatzer's size was chosen to be equal to or 1 mm less than the stretched diameter. The device was advanced transvenously into a 7F long guiding sheath and deployed under fluoroscopic and ultrasound guidance. Once its position was optimal, it was released. RESULTS The mean ASD diameter by transesophageal echocardiography was 14.1+/-2.3 mm and was significantly smaller (p < 0.001) than the stretched diameter of the ASD (16.8+/-2.4 mm). The mean device diameter was 16.6+/-2.3 mm. No complications were observed. After deployment of the prosthesis, there was no residual shunt in 13 (81.3%) of 16 patients. In three patients there was trivial residual shunt immediately after the procedure that had disappeared in two of them at the 3-month follow-up. CONCLUSIONS The Amplatzer is an efficient prosthesis that can be safely applied in children with secundum ASD. However, a study including a large number of patients and a longer follow-up period are required before this technique can be widely used.
منابع مشابه
Transcatheter closure of atrial septal defects--is balloon sizing still necessary?
INTRODUCTION The device closure of atrial septal defects has evolved over the years. In the early days of transcatheter occlusion, balloon sizing was used to choose an appropriate sized device. We postulate that balloon sizing does not value-add to the procedure and is unnecessary. MATERIALS AND METHODS Patients who had balloon sizing, with (Group 1, n = 38) or without (Group 2, n = 21) atria...
متن کاملAssessment of the geometric profile of the Amplatzer and Cardioseal septal occluders by three dimensional echocardiography.
OBJECTIVE To apply three dimensional echocardiography to describe the geometric profile of the Amplatzer and Cardioseal occluders after deployment for closure of atrial septal defect. METHODS 20 patients (mean (SD) age, 14 (5) years) were enrolled for transcatheter closure of a secundum atrial septal defect with the Amplatzer occluder (10) or with the Cardioseal occluder (10). The two populat...
متن کاملTranscatheter closure of secundum atrial septal defects and fenestrated Fontan using the Amplatzer septal occluder. Initial prospective study.
OBJECTIVE To evaluate the safety and efficacy of transcatheter closure of secundum atrial septal defects and fenestrated Fontan with the Amplatzer septal occluder. METHODS Fifteen consecutive patients, with a significant interatrial communications, were considered for the procedure; four patients with defects that were too large or with deficient margins were excluded after initial transesoph...
متن کاملDevice closure of atrial septal defect with amplatzer septal occluder in adults- safety and outcome.
BACKGROUND Secundum atrial septal defect is a common congenital heart disease. Amplatzer septal occluder has become the most commonly used device for its closure. This study was conducted to determine the safety and outcome of atrial septal defect closure with the Amplatzer septal occlude (ASO). METHODS This Case-series was conducted at Armed Forces Institute of Cardiology/National Institute ...
متن کاملSubacute, Silent Embolization of Amplatzer Atrial Septal Defect Closure Device to the Pulmonary Artery
Embolization of the closure device is a rare but potentially fatal complication of percutaneous atrial septal defect (ASD) closure. We report a case of 45-year-old woman who underwent ASD device closure with 32 mm Amplatzer device, which was embolized to the pulmonary artery without symptom one day after successful device implantation.
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of the American College of Cardiology
دوره 31 5 شماره
صفحات -
تاریخ انتشار 1998