Diagnostic performance and problem analysis of commercial tuberculosis antibody detection kits in China
نویسندگان
چکیده
BACKGROUND The diagnosis of bacterium-negative pulmonary tuberculosis (TB) and extra-pulmonary TB is challenging clinically. The detection of the anti-TB antibody has an important, auxiliary, clinical diagnostic value. Therefore, TB antibody detection kits should be screened and evaluated, and the reagents with the highest sensitivity and specificity should be chosen and used clinically. METHODS The diagnostic performance of 7 commercially available TB antibody detection kits (kits A, B, C, D, E, F and G) based on the gold immunoassay detection of immunoglobulin (Ig) G or IgM antibodies were simultaneously evaluated and compared in 62 TB cases and 56 non-TB cases in a laboratory. A retrospective analysis including 2549 cases was carried out to assess the clinical diagnosis values of bacteriological examinations and TB antibody tests (kits B and H used in the clinic). RESULTS The sensitivities of TB antibody kits A, B, C, D, E, F and G in the sera from 62 TB patients were 50.0%, 83.9%, 38.7%, 9.7%, 48.4%, 69.4% and 79.0%, respectively; the sensitivities in the sera from 24 smear-negative TB patients were 29.2%, 79.2%, 29.2%, 12.5%, 29.2%, 54.2% and 79.2%, respectively; the specificities in the sera from 56 non-TB patients were 73.2%, 25.0%, 85.7%, 96.4%, 78.6%, 78.6% and 50.0%, respectively. Of the 2549 clinically diagnosed cases, there were 1752 pulmonary TB cases, 505 extra-pulmonary TB cases, 87 old pulmonary TB cases and 205 non-TB cases. The positive results for smear, culture, TB antibody kit B and kit H in pulmonary TB cases were 39.8% (543/1365), 48.6% (372/765), 45.8% (802/1752) and 25.2% (442/1752), respectively; the results in extra-pulmonary TB cases were 3.4% (6/178), 5.8% (4/69), 35.4% (179/505), and 11.3% (57/505), respectively; the results in old pulmonary TB cases were 0% (0/64), 0% (0/30), 32.2% (28/87), and 9.2% (8/87), respectively; and the results in non-TB cases were 0% (0/121), 0% (0/56), 21.5% (44/205), and 2.4% (5/205), respectively. Of 624 smear-positive and/or culture-positive pulmonary TB cases, the sensitivities of antibody test kits B and H were 53.0% and 36.4%, respectively. Of 901 smear-negative and/or culture-negative pulmonary TB cases, the sensitivities of antibody test kits B and H were 42.5% and 19.0%, respectively. The positive rate of antibody detection in the bacterium-positive pulmonary TB cases was significantly higher than that in the bacterium-negative pulmonary TB cases (P < 0.05). CONCLUSIONS The colloidal gold-labeled TB antibody IgG detection assay is a simple, rapid and economical method that provides a better clinical auxiliary diagnosis value on TB, especially in smear-negative pulmonary TB and extra-pulmonary TB. The production, quality control, screening and evaluation of antibody detection kits are very important for its clinical application.
منابع مشابه
Letter to editor: comparative evaluation of four commercial serological antibody detection kits for the diagnosis of tuberculosis in China.
Rapid antibody detection has been used widely as a diagnostic aid in the diagnosis of tuberculosis (TB), especially for bacterium-negative TB. We evaluated the diagnostic value of four commercial antibody detection kits with 501 sera from TB patients and controls. The sensitivities and specificities of the four kits were as follows: (1) Rapid test A 71.8% and 91.4% (85.5% for non-TB diseases an...
متن کاملProduction of Erythropoietin-specific polyclonal Antibodies
Background: Erythropoietin, as a principal hormone promotes red blood cell production in bone marrow. Varieties of erythropoietin biosimilar are being produced by recombinant DNA technology in cell cultures. The detection or quantification of these molecules are being performed by diff erent methods which some of theme such as Western blot and enzymelinked immunosorbent assay (ELISA) require sp...
متن کاملEvaluation of in-house polymerase chain reaction assay sensitivity, can it be utilized in limited-resources settings?
Background :Polymerase chain reaction (PCR) assay has widely used for the detection of tuberculosis (TB). This study tried to compare in-house PCR with some well-known commercial PCR kits for detection of TB agent. Methods : Clinical samples obtained from 620 TB suspected patients were analyzed for the diagnosis of Mycobacterium tuberculosis complex (MTC) by in-house PCR. All samples were o...
متن کاملDevelopment of TaqMan Duplex Real-time PCR for Simultaneous Detection of Chlamydia Trachomatis and Mycoplasma Genitalium
Background and Objective: Sexually infections transmitted by bacteria are one of thetherapeutic and social problemsworldwide. The Real-time PCR assay is one of the most sensitive diagnostic and screening methods for these infections. The purpose of this study wassimultaneous detection of Chlamydia trachomatis and Mycoplasma genitaliumusing the TaqMan duplex real-time polymerase chain reaction. ...
متن کاملVariations in performance characteristics of commercial enzyme immunoassay kits for detection of antineutrophil cytoplasmic antibodies: what is the optimal cut off?
BACKGROUND Previous studies have shown considerable variation in diagnostic performance of enzyme linked immunosorbent assays (ELISAs) for measuring antineutrophil cytoplasmic antibodies (ANCA) specific for proteinase 3 (PR3) and myeloperoxidase (MPO). OBJECTIVE To analyse the performance characteristics of different commercially available direct ANCA ELISA kits. METHODS ELISA kits for dete...
متن کامل