Damage of an Oxinium femoral head and polyethylene liner following 'routine' total hip replacement.

نویسندگان

  • R W McCalden
  • K D Charron
  • R D Davidson
  • M G Teeter
  • D W Holdsworth
چکیده

We present a case of early retrieval of an Oxinium femoral head and corresponding polyethylene liner where there was significant surface damage to the head and polyethylene. The implants were retrieved at the time of revision surgery to correct leg-length discrepancy just 48 hours after the primary hip replacement. Appropriate analysis of the retrieved femoral head demonstrated loss of the Oxinium layer with exposure of the underlying substrate and transfer of titanium from the acetabular shell at the time of a reduction of the index total hip replacement. In addition, the level of damage to the polyethylene was extensive despite only 48 hours in situ. The purpose of this report is to highlight the care that is required at the time of reduction, especially with these hard femoral counter-faces such as Oxinium. To our knowledge, the damage occurring at the time of reduction has not been previously reported following the retrieval of an otherwise well-functioning hip replacement.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Surface damage to an Oxinium femoral head prosthesis after dislocation.

During open reduction of an irreducible anterior dislocation of a total hip replacement with an Oxinium femoral head, it was observed that the head had been significantly damaged. Gross and scanning electron microscopic examination revealed cracking, gouging, and delamination of the surface. Because of the risk which this poses for damaging the polyethylene acetabular liner, it is strongly reco...

متن کامل

The Effect of In Vivo Damage of Oxinium Femoral heads on the Wear of Highly Cross-linked Polyethylene

Bragdon, C.R; Wannomae, K. K.; Lozynsky, A.; Micheli, B.; Malchau, H. Harris Orthopaedic Biomechanics and Biomaterials Laboratory, Massachusetts General Hospital, Boston, MA 02114 Senior author [email protected] INTRODUCTION: Oxidized zirconium (OxiniumTM) is a bearing surface introduced for use in THA to minimize wear in vivo. Oxinium femoral heads are manufactured by oxidizing wrought Zr-...

متن کامل

Delamination of a highly cross-linked polyethylene liner associated with titanium deposits on the cobalt-chromium modular femoral head following dislocation.

Retrieval studies of total hip replacements with highly cross-linked ultra-high-molecular-weight polyethylene liners have shown much less surface damage than with conventional ultra-high-molecular-weight polyethylene liners. A recent revision hip replacement for recurrent dislocation undertaken after only five months revealed a highly cross-linked polyethylene liner with a large area of visible...

متن کامل

Highly Cross-linked Polyethylene Liner Dissociation from a Cement-less Modular Acetabular Shell: Two Case Reports

Liner dissociation of polyethylene from a cementless acetabular socket following total hip arthroplasty (THA) is a rare complication. Cross-linked polyethylene liner dissociation from AMS-HA shell (KYOCERA Med, Osaka, Japan) occurred in 2 out of the 4153 (0.04%) cases approximately 10 years after undergoing surgery at our institute. First case was an 80-year-old female who underwent right THA a...

متن کامل

Metallosis after Exchange of the Femoral Head and Liner following Ceramic Acetabular Liner Dissociation in Total Hip Arthroplasty with a Modular Layered Acetabular Component

The type of bearing material that should be used in revision surgery after the failure of ceramic-on-ceramic total hip arthroplasty (THA) remains controversial. In the case of ceramic fracture, the residual ceramic particles can cause consequent metallosis when metal implants are used for revision THA. On the other hand, in the case of THA failure without ceramic fracture, revision THA with a m...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 93 3  شماره 

صفحات  -

تاریخ انتشار 2011