Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy

نویسندگان

  • Nader Nassiri
  • Kourosh Sheibani
  • Sare Safi
  • Saman Nassiri
  • Alireza Ziaee
  • Farnaz Haji
  • Shiva Mehravaran
  • Nariman Nassiri
چکیده

PURPOSE To determine inter-device agreement for central corneal thickness (CCT) measurement among ultrasound pachymetry, rotating Scheimpflug imaging (Pentacam, Oculus, Wetzlar, Germany), and scanning slit corneal topography (Orbscan II, Bausch & Lomb, Rochester, NY, USA) in highly myopic eyes before and after photorefractive keratectomy (PRK). METHODS This prospective comparative study included 61 eyes of 32 patients with high myopia who underwent PRK. Six month postoperative CCT values were compared to preoperative values in 27 patients (51 eyes) who completed the follow up period. To determine the level of agreement, Pentacam and Orbscan II readings were compared to ultrasonic pachymetry measurements as the gold standard method. RESULTS Mean CCT measurements with ultrasound, Pentacam, and Orbscan II before PRK were 557µm, 556µm, and 564µm, respectively; and 451µm, 447µm, and 438µm 6 months after surgery in the same order. Preoperatively, the 95% limits of agreement (LoA) with ultrasound measurements were -20μm to 17μm for Pentacam and -21μm to 33μm for Orbscan II. Six months postoperatively, the 95% LoA were -30μm to 23μm for Pentacam and -69μm to 43μm for Orbscan II. CONCLUSION Preoperatively, CCT measurements were higher with Orbscan II as compared to ultrasound. Postoperatively, both Pentacam and Orbscan II measurements were lower than those obtained with ultrasound, but Pentacam had better agreement. The use of ultrasound, as the gold standard method, or Pentacam both appear to be preferable over Orbscan II among patients with high myopia.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy

PURPOSE The aim of this study was to determine the agreement between Pentacam HR (Scheimpflug imaging, Oculus) and Orbscan II (scanning slit topography, Bausch and Lomb) in measuring corneal parameters after photorefractive keratectomy (PRK) for hyperopia. METHODS In this prospective cross-sectional study, 38 hyperopic eyes undergoing PRK were examined before refractive surgery and 8 to 10 mo...

متن کامل

تغییرات برآمدگی سطح خلفی قرنیه پس از جراحی فتورفرکتیوکراتکتومی

  Background : Posterior corneal elevation changes are keys to early diagnosis of corneal ectasia after keratorefractive surgery, so it is important to use an accurate method for measuring posterior corneal elevation. The purpose of this study is investigating posterior corneal elevation changes with Pentacam(HR) on patients underwent photorefractive keratectomy.   Methods: In this clinical tri...

متن کامل

Central Corneal Thickness Measurement Using Ultrasonic Pachymetry, Rotating Scheimpflug Camera, and Scanning-slit Topography Exclusively in Thin Non-keratoconic Corneas

PURPOSE To evaluate the agreement among Pentacam, Orbscan and ultrasound (US) pachymetry for measurement of central corneal thickness (CCT) in thin corneas with normal topographic pattern. METHODS We included 88 eyes of 44 refractive surgery candidates with thinnest pachymetric readings of 500 micrometers (μm) or less on Orbscan, a normal topographic pattern, no sign of keratoconus, and best ...

متن کامل

Comparison of Central Corneal Thickness Measured through Pentacam, OCTCasia2, and Orbscan IIz in Normal Individuals

Purpose: To compare the central corneal thickness measured through Pentacam, OCTCasia2, and Orbscan IIz in normal individuals. Methods: In this cross sectional comparative study, the researcher investigated the eyes of 70 normal candidates (140 eyes) with the mean age of 31.42 ± 7.70, referring to Dr. Sedaghat eye clinic for refractive surgery. For each person, Central Corneal Thickness (CCT) w...

متن کامل

Central corneal thickness measurement with different devices in keratoconic patients

We read with great interest the article by Aghazadeh Amiri et al. concerning the corneal thickness (CT) measurements with Scheimpflug and slit scanning imaging techniques in keratoconus (KC). We would like to congratulate with the authors for their very interesting paper because it is well known that the measurement of CT is involved in the reliability of intraocular pressure measurements, both...