Relationship between Hounsfield Unit in CT Scan and Gray Scale in CBCT

نویسندگان

  • Tahmineh Razi
  • Mahdi Niknami
  • Fakhri Alavi Ghazani
چکیده

Background and aims. Cone-beam computed tomography (CBCT) is an imaging system which has many advantages over computed tomography (CT). In CT scan, Hounsfield Unit (HU) is proportional to the degree of x-ray attenuation by the tissue. In CBCT, the degree of x-ray attenuation is shown by gray scale (voxel value). The aim of the present study was to investigate the relationship between gray scale in CBCT) and Hounsfield Unit (HU) in CT scan. Materials and methods. In this descriptive study, the head of a sheep was scanned with 3 CBCT and one medical CT scanner. Gray scales and HUs were detected on images. Reconstructed data were analyzed to investigate relationship between CBCT gray scales and HUs. Results. A strong correlation between gray scales of CBCT and HUs of CT scan was determined. Conclusion. Considering the fact that gray scale in CBCT is the criteria in measurement of bone density before implant treatments, it is recommended because of the lower dose and cost compared to CT scan.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the gray level in CBCT systems and its relationship with HU in CT Scanners

Introduction: Cone-beam CT (CBCT) is an imaging system which offers three-dimensional (3D), multiplanar images and has many advantages over computed tomography (CT) including shorter acquisition times for the resolution desired in dentistry, lower radiation dose to the patient, reasonable price and higher spatial resolution but CBCT scanners are unable to display actual Hounsf...

متن کامل

In vitro Evaluation of the Relationship between Gray Scales in Digital Intraoral Radiographs and Hounsfield Units in CT Scans

Background: Jaw bone quality plays an essential role in treatment planning and prognosis of dental implants. Regarding several available methods for bone density measurements, they are not routinely used before implant surgery due to hard accessibility.Objective: An in vitro investigation of correlation between average gray scale in direct digital radiographs and Hounsfield units in CT-Scan pro...

متن کامل

Accuracy of densitometry of two cone beam computed tomography equipment in comparison with computed tomography

BACKGROUND The use of oral implants has been growing, and cone beam computerized tomography (CBCT) has become the method of choice for oral and maxillofacial radiology. OBJECTIVE To assess the accuracy of bone densitometry in two different CBCT devices in comparison with MDCT (multi-detector CT). METHODS Different concentrations of urografin, including 2.5%, 5%, 7.5%, 10%, 12.5%, were prepa...

متن کامل

In vitro Evaluation of the Relationship between Gray Scales in Digital Intraoral Radiographs and Hounsfield Units in CT Scans

BACKGROUND Jaw bone quality plays an essential role in treatment planning and prognosis of dental implants. Regarding several available methods for bone density measurements, they are not routinely used before implant surgery due to hard accessibility. OBJECTIVE An in vitro investigation of correlation between average gray scale in direct digital radiographs and Hounsfield units in CT-Scan pr...

متن کامل

Evaluation of gray scale changes of CBCT system images in different axis using the DICOM file

The images of dental CBCT imaging systems used in conic shaped beams, stored in the DICOM format, have various applications in the dentistry, including bone density estimation to select the location of the orthodontic implant, bone loss detection and etc. In these systems, unlike CT imaging systems, the resulting images exhibit gray-scale non-uniformity in each of the different axis in FOV. Thi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 8  شماره 

صفحات  -

تاریخ انتشار 2014