A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation
نویسندگان
چکیده
A 50-year-old man, who underwent a procedure for an implantable cardioverter defibrillator (ICD), visited the outpatient department of our clinic after suffering multiple ICD shocks. The ICD interrogation revealed recurrent shock due to a high frequency of noise that is sensed by the device as ventricular fibrillation. Chest radiography revealed a significant split in the insulation of the lead allowing the inner wire to protrude. We considered the removal of the failed lead, but the removal of ICD lead is potentially a high risk procedure, so we cut and capped a proximal part of the failed lead and inserted a new lead. This is the first report of a St. Jude Riata® dual coil defibrillator lead failure with clinical and radiologic evidence of a defect in lead insulation in Korea.
منابع مشابه
Thrombus formation on a defibrillator lead with conductor externalization
A 51 year old man with arrhythmogenic right ventricular cardiomyopathy reported non-specific symptoms of muscle aches and shivering at outpatient review. A dual chamber implantable cardioverter defibrillator (ICD) had been inserted seven years previously. Fluoroscopic lead screening had previously demonstrated conductor externalisation (Figure and Video) of a single coil Riata ICD lead (model n...
متن کاملImplantable cardioverter-defibrillator in a patient with dextrocardia situs inversus
Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...
متن کاملFluoroscopic investigation of Riata® transvenous defibrillator leads.
BACKGROUND Implantable cardioverter-defibrillator leads from Riata® family (St. Jude Medical Inc., Sylmar, CA, USA) have been recently recalled by Food and Drug Administration for concerns of a unique type of "inside-out" insulation failure leading to conductor externalization. The objective of this study was to evaluate the prevalence and predictors of conductor externalization in patients imp...
متن کاملA case of Durata ICD lead coil externalization: Inside-out lead abrasion?
The St Jude Medical Durata lead was produced in response to the recalled Riata and Riata ST leads in November 2011 after concerns of lead durability and inside-out abrasion. In order to protect against this susceptibility, the Riata ST Optim and Durata leads were coated with an abrasionresistant layer of silicone–polyurethane copolymer (Optim) tubing. We report a case of Durata lead extraction ...
متن کاملShock-induced right ventricular pacing failure caused by a short circuit: Uncommon but life-threatening complication of the Riata lead
Introduction The Riata and Riata ST implantable cardioverter-defibrillator (ICD) leads, a narrow caliber silicone lead without an outer jacket, have been shown to be prone to structural and electrical failure. While much interest has been shown in potentially lethal high-voltage (HV) failures caused by electrical short circuits due to a breach in the ethylene tetrafluoroethylene (ETFE) insulati...